Quality assessment of automatically planned O‐Ring linac SBRT plans for pelvic lymph node metastases, finding the optimal minimum target size by comparison with robotic SBRT

Date

Date

Date
2023
Journal Article
Published version
cris.lastimport.scopus2025-06-22T03:42:56Z
cris.lastimport.wos2025-07-28T01:35:36Z
cris.virtual.orcidhttps://orcid.org/0000-0002-6403-204X
cris.virtualsource.orcidb7123bc3-b618-49ad-ba72-a86accc77de6
dc.contributor.institutionUniversity of Zurich
dc.date.accessioned2023-12-15T12:09:15Z
dc.date.available2023-12-15T12:09:15Z
dc.date.issued2023-09-22
dc.description.abstract

PurposeThe purpose of this study is to assess the quality of automatic planned O‐Ring Halcyon linac SBRT plans for pelvic lymph node metastases and to establish an absolute PTV volume threshold as a plan quality prediction criterion. Compliance of the plans to institutional SBRT plan evaluation criteria and differences in plan quality and treatment delivery times between Halcyon Linac and CyberKnife robotic SBRT were evaluated.MethodsTwenty‐one CyberKnife treatment plans were replanned for Halcyon. Prescription doses range was 26–40 Gy in mean three fractions. The mean/median planning target volume was 4.0/3.6 cm$^{3}$. Institutional criteria for the plan evaluation were: New Conformity Index (NCI), Conformity Index (CI), Modified Gradient Index (MGI), selectivity index reciprocal (PIV/TV${PIV}$), and the target coverage by prescription isodose (%PIV). Statistical analysis based on the receiver operating characteristic (ROC) curve was used to determine a plan quality predictor threshold of the PTV volume. Comparative analysis of normal tissue complication probabilities (NTCP) was used to assess the risk of toxicity in healthy tissues.ResultsSeventy‐one percent (n = 15)/95% (n = 20) of Halcyon and 81% (n = 17)/100% (n = 21) of CK plans fulfilled all ideal/tolerance criteria. For PTVs above a found optimal threshold of 2.6 cm$^{3}$ (71%, n = 15), no statistically significant difference was observed between the CI, NCI, PIV/TV${PIV}$, and MGI indexes of both groups, while the coverage (%PIV) was statistically but not clinically significantly different between cohorts. Significantly shorter delivery times are expected with Halcyon. No significant differences in NTCP were observed.ConclusionAll but one automatically optimized Halcyon treatment plans demonstrated ideal or acceptable performance. PTV threshold of 2.6 cm$^{3}$ can be used as decision criteria in clinical settings. The results of our study demonstrated the promising performance of the Halcyon for pelvic SBRT, although plan‐specific QA is required to verify machine performance during plan delivery.

dc.identifier.doi10.1002/acm2.14143
dc.identifier.issn1526-9914
dc.identifier.scopus2-s2.0-85171847795
dc.identifier.urihttps://www.zora.uzh.ch/handle/20.500.14742/211845
dc.identifier.wos001068344100001
dc.language.isoeng
dc.subjectRadiology
dc.subjectNuclear Medicine and imaging
dc.subjectInstrumentation
dc.subjectRadiation
dc.subject.ddc530 Physics
dc.title

Quality assessment of automatically planned O‐Ring linac SBRT plans for pelvic lymph node metastases, finding the optimal minimum target size by comparison with robotic SBRT

dc.typearticle
dcterms.accessRightsinfo:eu-repo/semantics/openAccess
dcterms.bibliographicCitation.journaltitleJournal of Applied Clinical Medical Physics
dcterms.bibliographicCitation.number12
dcterms.bibliographicCitation.originalpublishernameThe American College of Medical Physics and American Institute of Physics
dcterms.bibliographicCitation.pagestarte14143
dcterms.bibliographicCitation.pmid37738649
dcterms.bibliographicCitation.volume24
dspace.entity.typePublicationen
uzh.contributor.affiliationUniversity of Zurich
uzh.contributor.affiliationRadiotherapy Hirslanden
uzh.contributor.affiliationUniversity of Zurich, Radiotherapy Hirslanden
uzh.contributor.authorHernández, Katerine Viviana Díaz
uzh.contributor.authorUnterkirhers, Sergejs
uzh.contributor.authorSchneider, Uwe
uzh.contributor.correspondenceYes
uzh.contributor.correspondenceNo
uzh.contributor.correspondenceNo
uzh.document.availabilitypublished_version
uzh.eprint.datestamp2023-12-15 12:09:15
uzh.eprint.lastmod2025-07-28 01:42:20
uzh.eprint.statusChange2023-12-15 12:09:15
uzh.harvester.ethYes
uzh.harvester.nbNo
uzh.identifier.doi10.5167/uzh-239224
uzh.jdb.eprintsId30230
uzh.oastatus.unpaywallgold
uzh.oastatus.zoraGold
uzh.publication.citationHernández, Katerine Viviana Díaz; Unterkirhers, Sergejs; Schneider, Uwe (2023). Quality assessment of automatically planned O‐Ring linac SBRT plans for pelvic lymph node metastases, finding the optimal minimum target size by comparison with robotic SBRT. Journal of Applied Clinical Medical Physics, 24(12):e14143.
uzh.publication.freeAccessAtdoi
uzh.publication.originalworkoriginal
uzh.publication.publishedStatusfinal
uzh.scopus.impact1
uzh.scopus.subjectsRadiation
uzh.scopus.subjectsInstrumentation
uzh.scopus.subjectsRadiology, Nuclear Medicine and Imaging
uzh.workflow.doajuzh.workflow.doaj.true
uzh.workflow.eprintid239224
uzh.workflow.fulltextStatuspublic
uzh.workflow.revisions43
uzh.workflow.rightsCheckkeininfo
uzh.workflow.sourceCrossref:10.1002/acm2.14143
uzh.workflow.statusarchive
uzh.wos.impact0
Files

Original bundle

Name:
J_Applied_Clin_Med_Phys___2023___Hern_ndez___Quality_assessment_of_automatically_planned_O_Ring_linac_SBRT_plans_for_pelvic.pdf
Size:
1.18 MB
Format:
Adobe Portable Document Format
Publication available in collections: