Publication:

Safety and feasibility of novel technology fusing echocardiography and fluoroscopy images during MitraClip interventions

Date

Date

Date
2014
Journal Article
Published version
cris.lastimport.scopus2025-07-29T03:39:23Z
cris.lastimport.wos2025-08-10T01:33:12Z
dc.contributor.institutionUniversity of Zurich
dc.date.accessioned2014-02-04T07:35:15Z
dc.date.available2014-02-04T07:35:15Z
dc.date.issued2014
dc.description.abstract

Aims: The EchoNavigator (EN) software (Philips Healthcare, Best, The Netherlands) enables real-time fusion of echocardiography and fluoroscopy by co-registration of the echocardiography probe on the x-ray image. We aimed to evaluate the feasibility and safety of this novel software during MitraClip procedures. Methods and results: Twenty-one patients were treated with the support of EchoNavigator software (EN+ patients). The primary (safety) endpoint was the total radiation dose. Secondary endpoints were fluoroscopy and total procedure time. The measurements were compared to those of 21 patients treated immediately before the installation of EchoNavigator (EN- patients). More MitraClips (45 vs. 36) were implanted in the EN+ group, mirroring more complex interventions in this group. In EN+ patients, radiation dose (Gy/cm(2)) was similar compared to EN- patients (146.5±123.6 vs.146.8±134.1, p=0.9). Total procedure time (minutes) was similar in the EN+ group compared to EN- patients (136.2±50.2 vs. 125.7±51.2, p=0.5). The main benefit of the EchoNavigator is the automated real-time fusion of echocardiography and fluoroscopy, leading to easier catheter manipulation. Conclusions: The use of EchoNavigator software was feasible and safe in all study patients. Further studies are necessary to confirm the benefits of using this software.

dc.identifier.doi10.4244/EIJV9I10A203
dc.identifier.issn1774-024X
dc.identifier.scopus2-s2.0-84894441534
dc.identifier.urihttps://www.zora.uzh.ch/handle/20.500.14742/100392
dc.identifier.wos000333444300014
dc.language.isoeng
dc.subject.ddc610 Medicine & health
dc.title

Safety and feasibility of novel technology fusing echocardiography and fluoroscopy images during MitraClip interventions

dc.typearticle
dcterms.accessRightsinfo:eu-repo/semantics/closedAccess
dcterms.bibliographicCitation.journaltitleEuroIntervention
dcterms.bibliographicCitation.number10
dcterms.bibliographicCitation.originalpublishernameEuropa Edition
dcterms.bibliographicCitation.pageend1216
dcterms.bibliographicCitation.pagestart1210
dcterms.bibliographicCitation.pmid24103772
dcterms.bibliographicCitation.urlhttp://www.pcronline.com/eurointervention/ahead_of_print/201310-02/
dcterms.bibliographicCitation.volume9
dspace.entity.typePublicationen
uzh.contributor.affiliationUniversitatsSpital Zurich
uzh.contributor.affiliationUniversitatsSpital Zurich
uzh.contributor.affiliationUniversitatsSpital Zurich
uzh.contributor.affiliationUniversitatsSpital Zurich
uzh.contributor.affiliationETH Zürich
uzh.contributor.affiliationETH Zürich
uzh.contributor.affiliationUniversitatsSpital Zurich
uzh.contributor.affiliationUniversitatsSpital Zurich
uzh.contributor.authorSündermann, Simon H
uzh.contributor.authorBiaggi, Patric
uzh.contributor.authorGrünenfelder, Jürg
uzh.contributor.authorGessat, Michael
uzh.contributor.authorFelix, Christian
uzh.contributor.authorBettex, Dominique
uzh.contributor.authorFalk, Volkmar
uzh.contributor.authorCorti, Roberto
uzh.contributor.correspondenceYes
uzh.contributor.correspondenceNo
uzh.contributor.correspondenceNo
uzh.contributor.correspondenceNo
uzh.contributor.correspondenceNo
uzh.contributor.correspondenceNo
uzh.contributor.correspondenceNo
uzh.contributor.correspondenceNo
uzh.document.availabilityno_document
uzh.eprint.datestamp2014-02-04 07:35:15
uzh.eprint.lastmod2025-08-10 01:54:50
uzh.eprint.statusChange2014-02-04 07:35:15
uzh.harvester.ethNo
uzh.harvester.nbNo
uzh.jdb.eprintsId27363
uzh.oastatus.unpaywallclosed
uzh.oastatus.zoraClosed
uzh.publication.citationSündermann, Simon H; Biaggi, Patric; Grünenfelder, Jürg; Gessat, Michael; Felix, Christian; Bettex, Dominique; Falk, Volkmar; Corti, Roberto (2014). Safety and feasibility of novel technology fusing echocardiography and fluoroscopy images during MitraClip interventions. EuroIntervention, 9(10):1210-1216.
uzh.publication.freeAccessAtdoi
uzh.publication.originalworkoriginal
uzh.publication.publishedStatusfinal
uzh.scopus.impact48
uzh.scopus.subjectsCardiology and Cardiovascular Medicine
uzh.workflow.eprintid90143
uzh.workflow.fulltextStatusnone
uzh.workflow.revisions55
uzh.workflow.rightsCheckkeininfo
uzh.workflow.statusarchive
uzh.wos.impact44
Publication available in collections: