Publication: Flow irregularities from syringe infusion pumps caused by syringe stiction
Flow irregularities from syringe infusion pumps caused by syringe stiction
Date
Date
Date
Citations
Peduzzi, N., Batliner, M., Grass, B., Buehler, P. K., Schmid Daners, M., Meboldt, M., & Weiss, M. (2020). Flow irregularities from syringe infusion pumps caused by syringe stiction. Paediatric Anaesthesia, 30(8), 885–891. https://doi.org/10.1111/pan.13926
Abstract
Abstract
Abstract
Objective: The current study aimed to evaluate the extent of the slide-stick phenomenon in differently designed infusion syringes at various infusion rates and filling positions.
Methods: Fluid delivery from three 50-mL infusion syringe brands (BD; Codan; Fresenius) was investigated using a flow sensor at flow rates of 0.5, 1.0, or 5.0 mL h-1 , with the syringes filled with either 10, 30, or 50 mL of distilled water. Two identical models (A/B) of the same infusion pump model were used. The effect of flow rate variations on the plasm
Metrics
Views
Additional indexing
Creators (Authors)
Volume
Volume
Volume
Number
Number
Number
Page range/Item number
Page range/Item number
Page range/Item number
Page end
Page end
Page end
Item Type
Item Type
Item Type
Dewey Decimal Classifikation
Dewey Decimal Classifikation
Dewey Decimal Classifikation
Keywords
Language
Language
Language
Publication date
Publication date
Publication date
Date available
Date available
Date available
ISSN or e-ISSN
ISSN or e-ISSN
ISSN or e-ISSN
OA Status
OA Status
OA Status
Publisher DOI
Metrics
Views
Citations
Peduzzi, N., Batliner, M., Grass, B., Buehler, P. K., Schmid Daners, M., Meboldt, M., & Weiss, M. (2020). Flow irregularities from syringe infusion pumps caused by syringe stiction. Paediatric Anaesthesia, 30(8), 885–891. https://doi.org/10.1111/pan.13926