Publication:

Characterization of spatial integrity with active and passive implants in a low-field magnetic resonance linear accelerator scanner

Date

Date

Date
2024
Journal Article
Published version
cris.lastimport.scopus2025-06-26T03:31:28Z
cris.lastimport.wos2025-07-30T01:30:32Z
dc.contributor.institutionUniversity of Zurich
dc.date.accessioned2024-05-06T14:47:12Z
dc.date.available2024-05-06T14:47:12Z
dc.date.issued2024-04-18
dc.description.abstract

BACKGROUND AND PURPOSE Standard imaging protocols can guarantee the spatial integrity of magnetic resonance (MR) images utilized in radiotherapy. However, the presence of metallic implants can significantly compromise this integrity. Our proposed method aims at characterizing the geometric distortions induced by both passive and active implants commonly encountered in planning images obtained from a low-field 0.35 T MR-linear accelerator (LINAC).

MATERIALS AND METHODS We designed a spatial integrity phantom defining 1276 control points and covering a field of view of 20x20x20 cm3. This phantom was scanned in a water tank with and without different implants used in hip and shoulder arthroplasty procedures as well as with active cardiac stimulators. The images were acquired with the clinical planning sequence (balanced steady-state free-precession, resolution 1.5x1.5x1.5 mm3). Spatial integrity was assessed by the Euclidian distance between the control point detected on the image and their theoretical locations. A first plane free of artefact (FPFA) was defined to evaluate the spatial integrity beyond the larger banding artefact.

RESULTS In the region extending up to 20 mm from the largest banding artefacts, the tested passive and active implants could cause distortions up to 2 mm and 3 mm, respectively. Beyond this region the spatial integrity was recovered and the image could be considered as unaffected by the implants.

CONCLUSIONS We characterized the impact of common implants on a low field MR-LINAC planning sequence. These measurements could support the creation of extra margin while contouring organs at risk and target volumes in the vicinity of implants.

dc.identifier.doi10.1016/j.phro.2024.100576
dc.identifier.issn2405-6316
dc.identifier.scopus2-s2.0-85190575663
dc.identifier.urihttps://www.zora.uzh.ch/handle/20.500.14742/219177
dc.identifier.wos001299948100001
dc.language.isoeng
dc.subject.ddc610 Medicine & health
dc.title

Characterization of spatial integrity with active and passive implants in a low-field magnetic resonance linear accelerator scanner

dc.typearticle
dcterms.accessRightsinfo:eu-repo/semantics/openAccess
dcterms.bibliographicCitation.journaltitlePhysics and Imaging in Radiation Oncology
dcterms.bibliographicCitation.originalpublishernameElsevier
dcterms.bibliographicCitation.pagestart100576
dcterms.bibliographicCitation.pmid38644933
dcterms.bibliographicCitation.volume30
dspace.entity.typePublicationen
uzh.contributor.affiliationUniversitatsSpital Zurich
uzh.contributor.affiliationUniversitatsSpital Zurich
uzh.contributor.affiliationUniversitatsSpital Zurich
uzh.contributor.affiliationUniversitatsSpital Zurich
uzh.contributor.affiliationUniversitatsSpital Zurich
uzh.contributor.affiliationUniversitatsSpital Zurich
uzh.contributor.affiliationUniversitatsSpital Zurich
uzh.contributor.affiliationUniversitatsSpital Zurich
uzh.contributor.authorPouymayou, Bertrand
uzh.contributor.authorPerez-Haas, Yoel
uzh.contributor.authorAllemann, Florin
uzh.contributor.authorSaguner, Ardan M
uzh.contributor.authorAndratschke, Nicolaus
uzh.contributor.authorGuckenberger, Matthias
uzh.contributor.authorTanadini-Lang, Stephanie
uzh.contributor.authorWilke, Lotte
uzh.contributor.correspondenceYes
uzh.contributor.correspondenceNo
uzh.contributor.correspondenceNo
uzh.contributor.correspondenceNo
uzh.contributor.correspondenceNo
uzh.contributor.correspondenceNo
uzh.contributor.correspondenceNo
uzh.contributor.correspondenceNo
uzh.document.availabilitypublished_version
uzh.eprint.datestamp2024-05-06 14:47:12
uzh.eprint.lastmod2025-07-30 01:35:21
uzh.eprint.statusChange2024-05-06 14:47:12
uzh.harvester.ethYes
uzh.harvester.nbNo
uzh.identifier.doi10.5167/uzh-259435
uzh.jdb.eprintsId45342
uzh.oastatus.unpaywallgold
uzh.oastatus.zoraGold
uzh.oatransformation.contractTRUE
uzh.oatransformation.contractDate01.01.2024-31.12.2024
uzh.oatransformation.contractIDElsevier2024
uzh.oatransformation.contractNameScienceDirect
uzh.oatransformation.contractURL
uzh.publication.citationPouymayou, Bertrand; Perez-Haas, Yoel; Allemann, Florin; Saguner, Ardan M; Andratschke, Nicolaus; Guckenberger, Matthias; Tanadini-Lang, Stephanie; Wilke, Lotte (2024). Characterization of spatial integrity with active and passive implants in a low-field magnetic resonance linear accelerator scanner. Physics and Imaging in Radiation Oncology, 30:100576.
uzh.publication.freeAccessAtdoi
uzh.publication.originalworkoriginal
uzh.publication.publishedStatusfinal
uzh.scopus.impact0
uzh.scopus.subjectsRadiation
uzh.scopus.subjectsOncology
uzh.scopus.subjectsRadiology, Nuclear Medicine and Imaging
uzh.workflow.doajuzh.workflow.doaj.true
uzh.workflow.eprintid259435
uzh.workflow.fulltextStatuspublic
uzh.workflow.revisions38
uzh.workflow.rightsCheckkeininfo
uzh.workflow.sourcePubMed:PMID:38644933
uzh.workflow.statusarchive
uzh.wos.impact0
Files

Original bundle

Name:
2024_Pouymayou_Phys_Imaging_Radiat_Oncol_Characterization_of_spatial_integrity_with_active.pdf
Size:
3.97 MB
Format:
Adobe Portable Document Format