Feasibility, safety, and short-term outcome of leadless ultrasound-based endocardial left ventricular resynchronization in heart failure patients: results of the wireless stimulation endocardially for CRT (WiSE-CRT) study


Auricchio, Angelo; Delnoy, Peter-Paul; Butter, Christian; Brachmann, Johannes; Van Erven, Lieselot; Spitzer, Stefan; Moccetti, Tiziano; Seifert, Martin; Markou, Thanasie; Laszo, Karolyi; Regoli, François (2014). Feasibility, safety, and short-term outcome of leadless ultrasound-based endocardial left ventricular resynchronization in heart failure patients: results of the wireless stimulation endocardially for CRT (WiSE-CRT) study. Europace, 16(5):681-688.

Abstract

AIMS Left ventricular (LV) endocardial pacing may address the limitations in the selection of an LV pacing site and provide improvements in cardiac resynchronization therapy (CRT) effectiveness. We report on the feasibility, the safety, and the short-term outcome of a leadless ultrasound-based technology for LV endocardial resynchronization in heart failure (HF) patients enroled into the Wireless Stimulation Endocardially for CRT (WiSE-CRT) study. METHODS AND RESULTS Seventeen HF patients were enroled and categorized as: (i) patients in whom attempted coronary sinus lead implantation for CRT had failed (n = 7); (ii) patients with a previously implanted CRT device, not responding to CRT (n = 2); and (iii) patients with previously implanted pacemakers or implantable cardioverter-defibrillator and meeting the standard indications for CRT (n = 8). System implantation was achieved in 13 patients (76.5%); mean R-wave amplitude was 5.6 ± 3.2 mV and the mean pacing threshold was 1.6 ± 1.0 V, respectively. In one patient, no sufficient pacing thresholds were found; in three patients pericardial effusion occurred. Biventricular pacing was recorded in 83% and 92% of the patients at 1 month and 6 months, respectively. QRS duration was shorter during biventricular pacing compared with right ventricular pacing at 1 month (-41 ms; P = 0.0002) and 6 months (-42 ms; P = 0.0011), respectively. At the 6-month follow-up, two-thirds of the patients had at least one functional class change. Left ventricular ejection fraction significantly increased (P < 0.01) by 6 points at the 6-month follow-up. CONCLUSION The feasibility of providing an endocardial stimulation for CRT with a leadless technology was successfully demonstrated. Despite the promising results for a novel technology, further study is required to definitively conclude the safety and the performance of the system. CLINICAL TRIAL REGISTRATION INFORMATION NCT01294527.

Abstract

AIMS Left ventricular (LV) endocardial pacing may address the limitations in the selection of an LV pacing site and provide improvements in cardiac resynchronization therapy (CRT) effectiveness. We report on the feasibility, the safety, and the short-term outcome of a leadless ultrasound-based technology for LV endocardial resynchronization in heart failure (HF) patients enroled into the Wireless Stimulation Endocardially for CRT (WiSE-CRT) study. METHODS AND RESULTS Seventeen HF patients were enroled and categorized as: (i) patients in whom attempted coronary sinus lead implantation for CRT had failed (n = 7); (ii) patients with a previously implanted CRT device, not responding to CRT (n = 2); and (iii) patients with previously implanted pacemakers or implantable cardioverter-defibrillator and meeting the standard indications for CRT (n = 8). System implantation was achieved in 13 patients (76.5%); mean R-wave amplitude was 5.6 ± 3.2 mV and the mean pacing threshold was 1.6 ± 1.0 V, respectively. In one patient, no sufficient pacing thresholds were found; in three patients pericardial effusion occurred. Biventricular pacing was recorded in 83% and 92% of the patients at 1 month and 6 months, respectively. QRS duration was shorter during biventricular pacing compared with right ventricular pacing at 1 month (-41 ms; P = 0.0002) and 6 months (-42 ms; P = 0.0011), respectively. At the 6-month follow-up, two-thirds of the patients had at least one functional class change. Left ventricular ejection fraction significantly increased (P < 0.01) by 6 points at the 6-month follow-up. CONCLUSION The feasibility of providing an endocardial stimulation for CRT with a leadless technology was successfully demonstrated. Despite the promising results for a novel technology, further study is required to definitively conclude the safety and the performance of the system. CLINICAL TRIAL REGISTRATION INFORMATION NCT01294527.

Statistics

Citations

Dimensions.ai Metrics
117 citations in Web of Science®
143 citations in Scopus®
Google Scholar™

Altmetrics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > Cardiocentro Ticino
Dewey Decimal Classification:610 Medicine & health
Scopus Subject Areas:Health Sciences > Cardiology and Cardiovascular Medicine
Health Sciences > Physiology (medical)
Language:English
Date:May 2014
Deposited On:12 Feb 2015 08:16
Last Modified:26 Jan 2022 05:09
Publisher:Oxford University Press
ISSN:1099-5129
OA Status:Closed
Publisher DOI:https://doi.org/10.1093/europace/eut435
PubMed ID:24497573
Full text not available from this repository.