Infection rate of emergency bolt-kit vs. non-emergency conventional implanted silver bearing external ventricular drainage catheters


Fichtner, Jens; Jilch, Astrid; Stieglitz, Lennart Henning; Beck, Jürgen; Raabe, Andreas; Z'Graggen, Werner J (2014). Infection rate of emergency bolt-kit vs. non-emergency conventional implanted silver bearing external ventricular drainage catheters. Clinical Neurology and Neurosurgery, 122:70-76.

Abstract

BACKGROUND Bolt-kit systems are increasingly used as an alternative to conventional external cerebrospinal fluid (CSF) drainage systems. Since 2009 we regularly utilize bolt-kit external ventricular drainage (EVD) systems with silver-bearing catheters inserted manually with a hand drill and skull screws for emergency ventriculostomy. For non-emergency situations, we use conventional ventriculostomy with subcutaneous tunneled silver-bearing catheters, performed in the operating room with a pneumatic drill. This retrospective analysis compared the two techniques in terms of infection rates.

METHODS 152 patients (aged 17-85 years, mean=55.4 years) were included in the final analysis; 95 received bolt-kit silver-bearing catheters and 57 received conventionally implanted silver-bearing catheters. The primary endpoint combined infection parameters: occurrence of positive CSF culture, colonization of catheter tips, or elevated CSF white blood cell counts (>4/μl). Secondary outcome parameters were presence of microorganisms in CSF or on catheter tips. Incidence of increased CSF cell counts and number of patients with catheter malposition were also compared.

RESULTS The primary outcome, defined as analysis of combined infection parameters (occurrence of either positive CSF culture, colonization of the catheter tips or raised CSF white blood cell counts >4/μl)was not significantly different between the groups (58.9% bolt-kit group vs. 63.2% conventionally implanted group, p=0.61, chi-square-test). The bolt-kit group was non-inferior and not superior to the conventional group (relative risk reduction of 6.7%; 90% confidence interval: -19.9% to 25.6%). Secondary outcomes showed no statistically significant difference in the incidence of microorganisms in CSF (2.1% bolt-kit vs. 5.3% conventionally implanted; p=0.30; chi-square-test).

CONCLUSIONS This analysis indicates that silver-bearing EVD catheters implanted with a bolt-kit system outside the operating room do not significantly elevate the risk of CSF infection as compared to conventional implant methods.

Abstract

BACKGROUND Bolt-kit systems are increasingly used as an alternative to conventional external cerebrospinal fluid (CSF) drainage systems. Since 2009 we regularly utilize bolt-kit external ventricular drainage (EVD) systems with silver-bearing catheters inserted manually with a hand drill and skull screws for emergency ventriculostomy. For non-emergency situations, we use conventional ventriculostomy with subcutaneous tunneled silver-bearing catheters, performed in the operating room with a pneumatic drill. This retrospective analysis compared the two techniques in terms of infection rates.

METHODS 152 patients (aged 17-85 years, mean=55.4 years) were included in the final analysis; 95 received bolt-kit silver-bearing catheters and 57 received conventionally implanted silver-bearing catheters. The primary endpoint combined infection parameters: occurrence of positive CSF culture, colonization of catheter tips, or elevated CSF white blood cell counts (>4/μl). Secondary outcome parameters were presence of microorganisms in CSF or on catheter tips. Incidence of increased CSF cell counts and number of patients with catheter malposition were also compared.

RESULTS The primary outcome, defined as analysis of combined infection parameters (occurrence of either positive CSF culture, colonization of the catheter tips or raised CSF white blood cell counts >4/μl)was not significantly different between the groups (58.9% bolt-kit group vs. 63.2% conventionally implanted group, p=0.61, chi-square-test). The bolt-kit group was non-inferior and not superior to the conventional group (relative risk reduction of 6.7%; 90% confidence interval: -19.9% to 25.6%). Secondary outcomes showed no statistically significant difference in the incidence of microorganisms in CSF (2.1% bolt-kit vs. 5.3% conventionally implanted; p=0.30; chi-square-test).

CONCLUSIONS This analysis indicates that silver-bearing EVD catheters implanted with a bolt-kit system outside the operating room do not significantly elevate the risk of CSF infection as compared to conventional implant methods.

Statistics

Citations

Dimensions.ai Metrics
10 citations in Web of Science®
10 citations in Scopus®
Google Scholar™

Altmetrics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Clinic for Neurosurgery
Dewey Decimal Classification:610 Medicine & health
Scopus Subject Areas:Health Sciences > Surgery
Health Sciences > Neurology (clinical)
Language:English
Date:1 July 2014
Deposited On:17 Feb 2015 17:04
Last Modified:26 Jan 2022 05:37
Publisher:Elsevier
ISSN:0303-8467
OA Status:Closed
Publisher DOI:https://doi.org/10.1016/j.clineuro.2014.04.018
PubMed ID:24908221
Full text not available from this repository.