Navigation auf zora.uzh.ch

Search ZORA

ZORA (Zurich Open Repository and Archive)

Embroidered electrode with silver/titanium coating for long-term ECG monitoring

Weder, Markus; Hegemann, Dirk; Amberg, Martin; Hess, Markus; Boesel, Luciano F; Abächerli, Roger; Meyer, Veronika R; Rossi, René M (2015). Embroidered electrode with silver/titanium coating for long-term ECG monitoring. Sensors, 15(1):1750-1759.

Abstract

For the long-time monitoring of electrocardiograms, electrodes must be skin-friendly and non-irritating, but in addition they must deliver leads without artifacts even if the skin is dry and the body is moving. Today's adhesive conducting gel electrodes are not suitable for such applications. We have developed an embroidered textile electrode from polyethylene terephthalate yarn which is plasma-coated with silver for electrical conductivity and with an ultra-thin titanium layer on top for passivation. Two of these electrodes are embedded into a breast belt. They are moisturized with a very low amount of water vapor from an integrated reservoir. The combination of silver, titanium and water vapor results in an excellent electrode chemistry. With this belt the long-time monitoring of electrocardiography (ECG) is possible at rest as well as when the patient is moving.

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Clinic for Thoracic Surgery
Dewey Decimal Classification:610 Medicine & health
Scopus Subject Areas:Physical Sciences > Analytical Chemistry
Life Sciences > Biochemistry
Physical Sciences > Atomic and Molecular Physics, and Optics
Physical Sciences > Instrumentation
Physical Sciences > Electrical and Electronic Engineering
Date:2015
Deposited On:30 Jan 2015 12:59
Last Modified:12 Mar 2025 02:41
Publisher:MDPI Publishing
ISSN:1424-8220
OA Status:Gold
Free access at:PubMed ID. An embargo period may apply.
Publisher DOI:https://doi.org/10.3390/s150101750
PubMed ID:25599424
Download PDF  'Embroidered electrode with silver/titanium coating for long-term ECG monitoring'.
Preview
  • Content: Published Version

Metadata Export

Statistics

Citations

Dimensions.ai Metrics
96 citations in Web of Science®
107 citations in Scopus®
Google Scholar™

Altmetrics

Downloads

160 downloads since deposited on 30 Jan 2015
22 downloads since 12 months
Detailed statistics

Authors, Affiliations, Collaborations

Similar Publications