Header

UZH-Logo

Maintenance Infos

Influence of airway-occluding instruments on airway pressure during jet ventilation for rigid bronchoscopy


Biro, P; Layer, M; Becker, H D; Herth, F; Wiedemann, K; Seifert, Burkhardt; Spahn, D R (2000). Influence of airway-occluding instruments on airway pressure during jet ventilation for rigid bronchoscopy. British Journal of Anaesthesia, 85(3):462-465.

Abstract

We measured changes in airway pressure (Paw) caused by microsurgical instruments introduced into a rigid bronchoscope during high frequency jet ventilation (HFJV). With approval of the institutional Ethics Committee, 10 adults undergoing elective tracheobronchial endoscopy and endosonography during general anaesthesia were investigated. Inflation of an endosonography probe balloon in the left main stem bronchus caused airway obstruction. Pressure measurements proximal and distal to the obstruction were compared after three degrees of obstruction (0%, 50% and 90%) and with two different driving pressure settings. Airway obstruction increased the mean (SD) peak inspiratory pressure (PIP) from 7.5 (2.6) to 9.5 (3.5) mm Hg for 2 atm (P = 0.0008) and from 9.7 (3.7) to 13.0 (5.1) mm Hg for 3 atm (P = 0.0001). Airway obstruction did not alter peripheral PIP (7.2 (4.1) to 7.1 (3.7) mm Hg for 2 atm and 8.8 (4.3) to 9.4 (5.2) mm for 3 atm), but resulted in an end-expiratory pressure (EEP) beyond the narrowing being significantly greater than in the unobstructed airway (2.5 (3.4) to 5.5 (3.7) mm Hg for 2 atm; P = 0.0005) and 3.2 (3.6) to 8.0 (4.3) mm for 3 atm; P < 0.0001). Severe airway narrowing increases inspiratory pressure proximal and expiratory pressure distal to the obstruction in relation to the applied driving pressure. Since the distal EEP never exceeded PIP, even near-total airway obstruction should not cause severe lung distension or barotrauma in subjects with normal lungs.

Abstract

We measured changes in airway pressure (Paw) caused by microsurgical instruments introduced into a rigid bronchoscope during high frequency jet ventilation (HFJV). With approval of the institutional Ethics Committee, 10 adults undergoing elective tracheobronchial endoscopy and endosonography during general anaesthesia were investigated. Inflation of an endosonography probe balloon in the left main stem bronchus caused airway obstruction. Pressure measurements proximal and distal to the obstruction were compared after three degrees of obstruction (0%, 50% and 90%) and with two different driving pressure settings. Airway obstruction increased the mean (SD) peak inspiratory pressure (PIP) from 7.5 (2.6) to 9.5 (3.5) mm Hg for 2 atm (P = 0.0008) and from 9.7 (3.7) to 13.0 (5.1) mm Hg for 3 atm (P = 0.0001). Airway obstruction did not alter peripheral PIP (7.2 (4.1) to 7.1 (3.7) mm Hg for 2 atm and 8.8 (4.3) to 9.4 (5.2) mm for 3 atm), but resulted in an end-expiratory pressure (EEP) beyond the narrowing being significantly greater than in the unobstructed airway (2.5 (3.4) to 5.5 (3.7) mm Hg for 2 atm; P = 0.0005) and 3.2 (3.6) to 8.0 (4.3) mm for 3 atm; P < 0.0001). Severe airway narrowing increases inspiratory pressure proximal and expiratory pressure distal to the obstruction in relation to the applied driving pressure. Since the distal EEP never exceeded PIP, even near-total airway obstruction should not cause severe lung distension or barotrauma in subjects with normal lungs.

Statistics

Citations

Dimensions.ai Metrics
16 citations in Web of Science®
16 citations in Scopus®
Google Scholar™

Altmetrics

Downloads

41 downloads since deposited on 15 Oct 2015
12 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > Epidemiology, Biostatistics and Prevention Institute (EBPI)
Dewey Decimal Classification:610 Medicine & health
Scopus Subject Areas:Health Sciences > Anesthesiology and Pain Medicine
Language:English
Date:September 2000
Deposited On:15 Oct 2015 07:56
Last Modified:30 Jul 2020 18:45
Publisher:Oxford University Press
ISSN:0007-0912
OA Status:Hybrid
Free access at:Publisher DOI. An embargo period may apply.
Publisher DOI:https://doi.org/10.1093/bja/85.3.462