Accuracy of non-invasive continuous total hemoglobin measurement by Pulse CO-Oximetry in severe traumatized and surgical bleeding patients


Baulig, Werner; Seifert, Burkhardt; Spahn, Donat R; Theusinger, Oliver M (2017). Accuracy of non-invasive continuous total hemoglobin measurement by Pulse CO-Oximetry in severe traumatized and surgical bleeding patients. Journal of Clinical Monitoring and Computing, 31(1):177-185.

Abstract

The Masimo Radical-7 Pulse CO-Oximeter (Masimo Corp., USA) non-invasively computes hemoglobin concentration (SpHb). SpHb was compared to Co-Oximeter readings (CoOxHb) of arterial samples in surgery patients of the emergency department. Forty-six patients were enrolled. The Masimo R1 25L (revision F and G) adult adhesive sensor was attached to the ring finger of the arterially cannulated hand. Before start, every 30 min during surgery and in the case of severe bleeding SpHb and CoOxHb values were documented. SpHb and post hoc adjusted SpHb (AdSpHb) values were analyzed. Linear regression analysis and Bland-Altman plot for agreement were performed. The detection failure rate of SpHb was 24.5 %. CoOxHb and SpHb showed a strong correlation (r = +0.81), but agreement was moderate [bias (LOA) of -0.6 (-3.0; +1.9)] g/dl. Positive and negative predicted value was 0.49 and 0.69. Exclusion of changes of CoOxHb values ≤1 g/dl resulted in a positive and negative predictive value of 0.66 and 1.00. Post hoc adjustment of the SpHb (AdSpHb) improved linear correlation of CoOxHb and AdSpHb [r = +0.90 (p < 0.001)] but less the agreement [bias (LOA) of CoOxHb and AdSpHb = -0.1 (-2.1/+1.9) g/dl]. SpHb agreed only moderately with CoOxHb values and predicted decreases of CoOxHb only if changes of SpHb ≤ 1.0 g/dl were excluded. The detection failure rate of SpHb was high. At present, additional refinements of the current technology are necessary to further improve performance of non-invasive hemoglobin measurement in the clinical setting.

Abstract

The Masimo Radical-7 Pulse CO-Oximeter (Masimo Corp., USA) non-invasively computes hemoglobin concentration (SpHb). SpHb was compared to Co-Oximeter readings (CoOxHb) of arterial samples in surgery patients of the emergency department. Forty-six patients were enrolled. The Masimo R1 25L (revision F and G) adult adhesive sensor was attached to the ring finger of the arterially cannulated hand. Before start, every 30 min during surgery and in the case of severe bleeding SpHb and CoOxHb values were documented. SpHb and post hoc adjusted SpHb (AdSpHb) values were analyzed. Linear regression analysis and Bland-Altman plot for agreement were performed. The detection failure rate of SpHb was 24.5 %. CoOxHb and SpHb showed a strong correlation (r = +0.81), but agreement was moderate [bias (LOA) of -0.6 (-3.0; +1.9)] g/dl. Positive and negative predicted value was 0.49 and 0.69. Exclusion of changes of CoOxHb values ≤1 g/dl resulted in a positive and negative predictive value of 0.66 and 1.00. Post hoc adjustment of the SpHb (AdSpHb) improved linear correlation of CoOxHb and AdSpHb [r = +0.90 (p < 0.001)] but less the agreement [bias (LOA) of CoOxHb and AdSpHb = -0.1 (-2.1/+1.9) g/dl]. SpHb agreed only moderately with CoOxHb values and predicted decreases of CoOxHb only if changes of SpHb ≤ 1.0 g/dl were excluded. The detection failure rate of SpHb was high. At present, additional refinements of the current technology are necessary to further improve performance of non-invasive hemoglobin measurement in the clinical setting.

Statistics

Citations

Dimensions.ai Metrics
16 citations in Web of Science®
17 citations in Scopus®
Google Scholar™

Altmetrics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > Epidemiology, Biostatistics and Prevention Institute (EBPI)
04 Faculty of Medicine > University Hospital Zurich > Institute of Anesthesiology
Dewey Decimal Classification:610 Medicine & health
Scopus Subject Areas:Health Sciences > Health Informatics
Health Sciences > Critical Care and Intensive Care Medicine
Health Sciences > Anesthesiology and Pain Medicine
Language:English
Date:2017
Deposited On:12 Jan 2016 10:48
Last Modified:08 Jul 2022 12:59
Publisher:Springer
ISSN:1387-1307
OA Status:Closed
Publisher DOI:https://doi.org/10.1007/s10877-015-9816-2
PubMed ID:26686689
Full text not available from this repository.