Evaluation of agreement and trending ability between transpulmonary thermodilution and calibrated pulse contour and pulse power cardiac output monitoring methods against pulmonary artery thermodilution in anesthetized dogs


Kutter, Annette P N; Bettschart-Wolfensberger, Regula; Romagnoli, Noemi; Bektas, Rima N (2016). Evaluation of agreement and trending ability between transpulmonary thermodilution and calibrated pulse contour and pulse power cardiac output monitoring methods against pulmonary artery thermodilution in anesthetized dogs. Journal of Veterinary Emergency and Critical Care, 26(4):531-540.

Abstract

Objective: To assess agreement and trending ability of transpulmonary thermodilution (TPTD), calibrated pulse contour (PiCCO), and pulse power (PulseCO) methods compared to pulmonary artery thermodilution (PATD) for determination of cardiac output (CO) in anesthetized dogs.
Design: Experimental, prospective study.
Setting: University teaching hospital.
Animals: Six adult Beagle dogs.
Interventions: Dogs were anesthetized with sevoflurane and instrumented with pulmonary and femoral artery thermodilution catheters. CO was measured at baseline and at 5, 15, 30, 45, 60, 120, 180, and 240 minutes after IV administration of ketamine or s-ketamine. Baseline PATD and TPTD calibrated PulseCO and PiCCO, respectively. Agreement and trending ability was analyzed with Bland–Altman, concordance, and polar plot methodology.
Measurements and Main Results: Median (range) CO values of 2.27 (0.98–3.4) L/min were measured with PATD, and 2.8 (1.9–4.04) L/min with TPTD, which resulted in a mean bias (± standard deviation) of −0.66 (± 0.36) L/min. Concordance rate was 91% and radial limits of agreement (RLOA) were ±35°. PATD against PiCCO resulted in a mean bias of −0.71 (± 0.62) L/min and PATD against PulseCO in a mean bias of 0.13 (± 0.46) L/min. The continuous techniques resulted in concordance rates of 77% for PATD-PiCCO and 74% for PATD-PulseCO and RLOA of ±57° and ±60°, respectively.
Conclusions: Intermittent TPTD showed marginal trending ability, while continuous pulse contour and pulse power methods showed poor trending ability over a 4-hour period. The poor performance and possible side effects of the methods tested in this study suggest that they should not be recommended for use in critical patients.

Abstract

Objective: To assess agreement and trending ability of transpulmonary thermodilution (TPTD), calibrated pulse contour (PiCCO), and pulse power (PulseCO) methods compared to pulmonary artery thermodilution (PATD) for determination of cardiac output (CO) in anesthetized dogs.
Design: Experimental, prospective study.
Setting: University teaching hospital.
Animals: Six adult Beagle dogs.
Interventions: Dogs were anesthetized with sevoflurane and instrumented with pulmonary and femoral artery thermodilution catheters. CO was measured at baseline and at 5, 15, 30, 45, 60, 120, 180, and 240 minutes after IV administration of ketamine or s-ketamine. Baseline PATD and TPTD calibrated PulseCO and PiCCO, respectively. Agreement and trending ability was analyzed with Bland–Altman, concordance, and polar plot methodology.
Measurements and Main Results: Median (range) CO values of 2.27 (0.98–3.4) L/min were measured with PATD, and 2.8 (1.9–4.04) L/min with TPTD, which resulted in a mean bias (± standard deviation) of −0.66 (± 0.36) L/min. Concordance rate was 91% and radial limits of agreement (RLOA) were ±35°. PATD against PiCCO resulted in a mean bias of −0.71 (± 0.62) L/min and PATD against PulseCO in a mean bias of 0.13 (± 0.46) L/min. The continuous techniques resulted in concordance rates of 77% for PATD-PiCCO and 74% for PATD-PulseCO and RLOA of ±57° and ±60°, respectively.
Conclusions: Intermittent TPTD showed marginal trending ability, while continuous pulse contour and pulse power methods showed poor trending ability over a 4-hour period. The poor performance and possible side effects of the methods tested in this study suggest that they should not be recommended for use in critical patients.

Statistics

Citations

Dimensions.ai Metrics
7 citations in Web of Science®
8 citations in Scopus®
Google Scholar™

Altmetrics

Downloads

2 downloads since deposited on 01 Feb 2016
1 download since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:05 Vetsuisse Faculty > Veterinary Clinic > Department of Clinical Diagnostics and Services
Dewey Decimal Classification:570 Life sciences; biology
630 Agriculture
Scopus Subject Areas:Health Sciences > General Veterinary
Uncontrolled Keywords:cardiac output, dog, pulse contour analysis, pulse power analysis, transpulmonary thermodilution
Language:English
Date:2016
Deposited On:01 Feb 2016 11:37
Last Modified:19 Mar 2023 15:58
Publisher:Wiley-Blackwell Publishing, Inc.
ISSN:1476-4431
OA Status:Closed
Publisher DOI:https://doi.org/10.1111/vec.12439
PubMed ID:26754858