PGA (polyglycolic acid)-P4HB (poly-4-hydroxybutyrate)-based bioengineered valves in the rat aortic circulation

Książek, Agnieszka A; Mitchell, Katharyn J; Cesarovic, Nikola; Schwarzwald, Colin C; Hoerstrup, Simon P; Weber, Benedikt (2016). PGA (polyglycolic acid)-P4HB (poly-4-hydroxybutyrate)-based bioengineered valves in the rat aortic circulation. Journal of Heart Valve Disease, 25(3):380-388.

Abstract

BACKGROUND: Bioengineered living autologous valves with remodeling and growth capacity represent a promising concept for future cardiac and venous valve repair. A meticulous understanding of the mechanisms involved in recellularization and remodeling is essential for the safe and efficient clinical translation of this technology. In this context, the first investigations of bioengineered vascular grafts in immune-incompetent or transgenic rodents represented an important step. However, the in-vivo assessment of bioengineered synthetic scaffold-based (biodegradable) valve replacements in rodent models has not been achieved to date.
METHODS: Miniaturized monocuspid PGA (polyglycolic acid)-P4HB (poly-4-hydroxybutyrate)-based valves were created, incorporated into metallic stents (length 2.0 mm, diameter 1.1 mm) and introduced into catheter-based implantation devices. Wistar outbred rats (n = 8) underwent a laparotomy, abdominal aorta arteriotomy and valve delivery into the abdominal aorta. Valve placement and function were evaluated following deployment using ultrasound (Doppler- and M-mode). Explanted tissues were analyzed both macroscopically and histopathologically.
RESULTS: No significant physiological or hemodynamic changes were observed, including heart rate, pressure gradients, velocity values and cardiac output before and after valve implantation. The cross-sectional area at the level of the stented valve was reduced by 22%. Valvular leaflet oscillation was observed in two animals, and thrombus formation in the stent was observed in one animal. Histological evaluation revealed cellular infiltration within 3 h in vivo, and no signs of thrombus deposition on the valvular surface.
CONCLUSIONS: This study demonstrated the technical feasibility of the transcatheter implantation of bioengineered stented miniaturized valves into the infrarenal rat aorta, without affecting the animal's physiological and hemodynamic variables and with valvular oscillation in part of the implants. These results could serve as a basis for the implementation of a chronic rat in-vivo model for mechanistic studies in bioengineered valvular tissues under systemic hemodynamic conditions. Video 1: 2D ultrasonographic projection revealing graft's leaflet oscillation.

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Clinic for Cardiac Surgery
05 Vetsuisse Faculty > Veterinary Clinic > Equine Department
05 Vetsuisse Faculty > Center for Applied Biotechnology and Molecular Medicine
04 Faculty of Medicine > Institute for Regenerative Medicine (IREM)
Dewey Decimal Classification:570 Life sciences; biology
630 Agriculture
Scopus Subject Areas:Health Sciences > Cardiology and Cardiovascular Medicine
Language:English
Date:May 2016
Deposited On:26 Jan 2017 08:47
Last Modified:16 Nov 2024 04:34
Publisher:ICR Publishers
ISSN:0966-8519
OA Status:Closed
PubMed ID:27989051
Full text not available from this repository.

Metadata Export

Statistics

Citations

Dimensions.ai Metrics

Altmetrics

Authors, Affiliations, Collaborations

Similar Publications