Ultrasound needle visibility in contrast mode imaging: an in vitro and ex vivo study


Rominger, Marga B; Martini, Katharina; Dappa, Evelyn; Puippe, Gilbert Dominique; Klingmüller, Volker; Frauenfelder, Thomas; Sanabria, Sergio J (2017). Ultrasound needle visibility in contrast mode imaging: an in vitro and ex vivo study. Ultrasound international open, 3(2):E82-E88.

Abstract

PURPOSE: To evaluate needle visibility in ultrasound under contrast mode conditions.
MATERIALS AND METHODS: Needle visibility was evaluated for bevel, EchoTip (®) and shaft of 18G Chiba biopsy needle with a 9 MHz linear probe (GE Logiq E9). Insertion angles varied between 30°(steep) and 90°(parallel to the probe surface). The acoustic output varied from 5-28%. 2 different contrast mode presets with either 'Amplitude Modulation' (Penetration) or 'Phase Inversion Harmonics' (High Resolution) were assessed. All other imaging parameters were kept constant. The visibility of bevel, EchoTip (®) and shaft was assessed for grayscale and color-coded images with a 3-point Likert-like scale (not, poorly, well visible) by 2 independent readers. The echogenicity of the needle bevel, EchoTip (®) and shaft was assessed in deciBel (dB) on the color-coded images.
RESULTS: With the parallel insertion angle, all needle areas were well visible. With steep insertion the EchoTip (®) was the only visible area. High Resolution was superior to Penetration (p<0.001). The visibility and echogenicity of the needle bevel ( rgrayscale =0.109, p grayscale =0.178; rcolor-coded =0.236, p color-coded =0.266; rdB =0.956, p dB =0.001), EchoTip (®) ( rgrayscale += 0.477, p grayscale += 0.018; rcolor-coded =0.540, p color-coded += 0.006; rdB =0.911, p dB =0.001) and shaft ( rgrayscale =0.124, p grayscale =0.563; rcolor-coded =0.061, p color-coded += 0.775; rdB += 0.926, p dB =0.001) increased with increasing acoustic output. Grayscale images were superior to color-coded images for needle visibility (p=0.004).
CONCLUSION: Parallel needle insertion, use of an echogenic tip, adequate choice of presets, increased acoustic output, and dual view of grayscale and color-coded images improve needle visibility in ultrasound under contrast mode conditions.

Abstract

PURPOSE: To evaluate needle visibility in ultrasound under contrast mode conditions.
MATERIALS AND METHODS: Needle visibility was evaluated for bevel, EchoTip (®) and shaft of 18G Chiba biopsy needle with a 9 MHz linear probe (GE Logiq E9). Insertion angles varied between 30°(steep) and 90°(parallel to the probe surface). The acoustic output varied from 5-28%. 2 different contrast mode presets with either 'Amplitude Modulation' (Penetration) or 'Phase Inversion Harmonics' (High Resolution) were assessed. All other imaging parameters were kept constant. The visibility of bevel, EchoTip (®) and shaft was assessed for grayscale and color-coded images with a 3-point Likert-like scale (not, poorly, well visible) by 2 independent readers. The echogenicity of the needle bevel, EchoTip (®) and shaft was assessed in deciBel (dB) on the color-coded images.
RESULTS: With the parallel insertion angle, all needle areas were well visible. With steep insertion the EchoTip (®) was the only visible area. High Resolution was superior to Penetration (p<0.001). The visibility and echogenicity of the needle bevel ( rgrayscale =0.109, p grayscale =0.178; rcolor-coded =0.236, p color-coded =0.266; rdB =0.956, p dB =0.001), EchoTip (®) ( rgrayscale += 0.477, p grayscale += 0.018; rcolor-coded =0.540, p color-coded += 0.006; rdB =0.911, p dB =0.001) and shaft ( rgrayscale =0.124, p grayscale =0.563; rcolor-coded =0.061, p color-coded += 0.775; rdB += 0.926, p dB =0.001) increased with increasing acoustic output. Grayscale images were superior to color-coded images for needle visibility (p=0.004).
CONCLUSION: Parallel needle insertion, use of an echogenic tip, adequate choice of presets, increased acoustic output, and dual view of grayscale and color-coded images improve needle visibility in ultrasound under contrast mode conditions.

Statistics

Citations

Altmetrics

Downloads

66 downloads since deposited on 30 Aug 2017
65 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Clinic for Diagnostic and Interventional Radiology
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:April 2017
Deposited On:30 Aug 2017 17:19
Last Modified:19 Feb 2018 08:35
Publisher:Georg Thieme Verlag
ISSN:2199-7152
OA Status:Gold
Free access at:PubMed ID. An embargo period may apply.
Publisher DOI:https://doi.org/10.1055/s-0043-101511
PubMed ID:28653045

Download

Download PDF  'Ultrasound needle visibility in contrast mode imaging: an in vitro and ex vivo study'.
Preview
Content: Published Version
Language: English
Filetype: PDF
Size: 880kB
View at publisher
Licence: Creative Commons: Attribution-NonCommercial-NoDerivatives 4.0 International (CC BY-NC-ND 4.0)