Navigation auf zora.uzh.ch

Search

ZORA (Zurich Open Repository and Archive)

Comparison of procedural and clinical outcomes with evolut R versus medtronic corevalve a Swiss TAVI registry analysis

Noble, Stephane; Stortecky, Stefan; Heg, Dik; Tueller, David; Jeger, Raban; Toggweiler, Stefan; Ferrari, Enrico; Nietlispach, Fabian; Taramasso, Maurizio; Maisano, Francesco; Grünenfelder, Jürg; Jüni, Peter; Huber, Christoph; Carrel, Thierry; Windecker, Stephan; Wenaweser, Peter; Roffi, Marco (2017). Comparison of procedural and clinical outcomes with evolut R versus medtronic corevalve a Swiss TAVI registry analysis. EuroIntervention, 12(18):e2170-e2176.

Abstract

AIMS: Data on procedural and clinical outcomes after transcatheter aortic valve implantation (TAVI) with the new-generation self-expanding Medtronic Evolut R prosthesis in comparison with its predecessor, the Medtronic CoreValve, are scarce. The aim of this study was to assess the safety and efficacy of the Evolut R device compared with the former-generation CoreValve.
METHODS AND RESULTS: In a nationwide, prospective, multicentre cohort study, outcomes of consecutive transfemoral TAVI patients treated with the new-generation Medtronic Evolut R (September 2014 - February 2016) and the Medtronic CoreValve (February 2011 - February 2016) were investigated. Events were reported according to VARC-2 and adjudicated by a clinical events committee. During the study period, 317 and 678 consecutive patients underwent TAVI with the Evolut R and the CoreValve bioprosthesis, respectively. Baseline clinical characteristics between the groups were comparable, although Evolut R patients were lower risk according to the STS score (4.8±3.4% vs. 6.9±5.0%, p<0.001) and logistic EuroSCORE (17.3±13% vs. 20.1±13%, p=0.009). Implantation of the Evolut R was associated with a lower use of predilatation (48.1% vs. 72.4%, p<0.001), a shorter procedure time (67.9±36 min vs. 76.7±42 min, p=0.002), and less contrast dye use during the procedure (155.2±98 ml vs. 208.0±117 ml, p<0.001). Post-procedural mean gradient was comparable (7.4±4.7 mmHg vs. 7.5±5.0 mmHg), as were the 30-day rates of moderate to severe aortic regurgitation (8.5% vs. 10.5%), major vascular (9.8% vs. 10.3%) and life-threatening bleeding complications (5.4% vs. 5.3%), disabling stroke (1.9% vs. 1.6%), all-cause mortality (3.2% vs. 3.4%) as well as permanent pacemaker implantation (22.1% vs. 23.4%).
CONCLUSIONS: Thirty-day clinical outcomes were favourable and comparable between the Evolut R and the CoreValve bioprosthesis.

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Clinic for Cardiac Surgery
04 Faculty of Medicine > University Hospital Zurich > Clinic for Cardiology
04 Faculty of Medicine > Cardiocentro Ticino
Dewey Decimal Classification:610 Medicine & health
Scopus Subject Areas:Health Sciences > Cardiology and Cardiovascular Medicine
Language:English
Date:7 April 2017
Deposited On:18 Dec 2017 15:24
Last Modified:20 Aug 2024 03:41
Publisher:Europa Digital and Publishing
ISSN:1774-024X
OA Status:Closed
Free access at:Publisher DOI. An embargo period may apply.
Publisher DOI:https://doi.org/10.4244/EIJ-D-16-00677
PubMed ID:28067197