Navigation auf zora.uzh.ch

Search

ZORA (Zurich Open Repository and Archive)

Analysis of the precision of variable flip angle T1 mapping with emphasis on the noise propagated from RF transmit field maps

Lee, Yoojin; Callaghan, Martina F; Nagy, Zoltán (2017). Analysis of the precision of variable flip angle T1 mapping with emphasis on the noise propagated from RF transmit field maps. Frontiers in Neuroscience, 11:106.

Abstract

In magnetic resonance imaging, precise measurements of longitudinal relaxation time (T1) is crucial to acquire useful information that is applicable to numerous clinical and neuroscience applications. In this work, we investigated the precision of T1 relaxation time as measured using the variable flip angle method with emphasis on the noise propagated from radiofrequency transmit field (B+1) measurements. The analytical solution for T1 precision was derived by standard error propagation methods incorporating the noise from the three input sources: two spoiled gradient echo (SPGR) images and a B+1 map. Repeated in vivo experiments were performed to estimate the total variance in T1 maps and we compared these experimentally obtained values with the theoretical predictions to validate the established theoretical framework. Both the analytical and experimental results showed that variance in the B+1 map propagated comparable noise levels into the T1 maps as either of the two SPGR images. Improving precision of the B+1 measurements significantly reduced the variance in the estimated T1 map. The variance estimated from the repeatedly measured in vivo T1 maps agreed well with the theoretically-calculated variance in T1 estimates, thus validating the analytical framework for realistic in vivo experiments. We concluded that for T1 mapping experiments, the error propagated from the B+1 map must be considered. Optimizing the SPGR signals while neglecting to improve the precision of the B+1 map may result in grossly overestimating the precision of the estimated T1 values.

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:03 Faculty of Economics > Department of Economics
Dewey Decimal Classification:330 Economics
Scopus Subject Areas:Life Sciences > General Neuroscience
Uncontrolled Keywords:B1+, map, T1 map, error propagation, uncertainty, precision, variable flip angle
Scope:Discipline-based scholarship (basic research)
Language:English
Date:March 2017
Deposited On:23 Jan 2018 19:29
Last Modified:22 Aug 2024 03:33
Publisher:Frontiers Research Foundation
ISSN:1662-453X
OA Status:Gold
Free access at:PubMed ID. An embargo period may apply.
Publisher DOI:https://doi.org/10.3389/fnins.2017.00106
PubMed ID:28337119
Other Identification Number:merlin-id:15861
Project Information:
  • Funder: SNSF
  • Grant ID: 31003A_166118
  • Project Title: Advanced non-invasive human in-vivo cortical parcellation using multimodal MRI
Download PDF  'Analysis of the precision of variable flip angle T1 mapping with emphasis on the noise propagated from RF transmit field maps'.
Preview
  • Content: Published Version
  • Licence: Creative Commons: Attribution 4.0 International (CC BY 4.0)

Metadata Export

Statistics

Citations

Dimensions.ai Metrics
18 citations in Web of Science®
20 citations in Scopus®
Google Scholar™

Altmetrics

Downloads

208 downloads since deposited on 23 Jan 2018
9 downloads since 12 months
Detailed statistics

Authors, Affiliations, Collaborations

Similar Publications