Header

UZH-Logo

Maintenance Infos

Structural, morphological, compositional, and mechanical changes of palatal implants after use: a retrieval analysis


Schätzle, Marc; Zinelis, Spiros; Markic, Goran; Eliades, George; Eliades, Theodore (2017). Structural, morphological, compositional, and mechanical changes of palatal implants after use: a retrieval analysis. European Journal of Orthodontics, 39(6):579-585.

Abstract

Purpose The aim of this study was to characterize the surface, elemental, and mechanical alterations of orthodontic palatal implants after intraoral aging. Materials and method Nineteen consecutively retrieved implants (RET) after orthodontic treatment and three unused implants used as control (CON) were included in this study. Both groups were characterized non-destructively by Stereomicroscopy, Optical Profilometry (Sa, Sq, Sz, Sc), and SEM/EDX analysis and then destructively after metallogaphic preparation employing instrumented indentation testing (HM, EIT, ηIT, and HV) and SEM/EDX at bone-implant interface. Results All retrieved implants showed a loss of gloss with the formation of bone-like formation on the majority of them. However, no differences in surface roughness parameters were identified between macroscopically intact and retrieved regions of implants. The elements precipitated on the surface were O, C, Ca, and P while traces of Na, K, Al, S, Cl, and Mg were also identified. The surface of control sample is characterized by small pits while only Ti and Al traces were identified by EDX analysis. The presence of all the aforementioned elements apart from Ti and Al on the retrieved implants' surface should be appended to the contact of implant with bone and biological fluids while Interfacial analysis revealed a well-formed bone-implant interface. However, no significant differences were found for all mechanical properties tested between RET and CON groups. Conclusions The results of this study indicate that retrieved palatal implant surface has undergone morphological and elemental alterations probably associated with the osseointegration process during service. Insertion and functional loading did not affect the mechanical properties of implants tested.

Abstract

Purpose The aim of this study was to characterize the surface, elemental, and mechanical alterations of orthodontic palatal implants after intraoral aging. Materials and method Nineteen consecutively retrieved implants (RET) after orthodontic treatment and three unused implants used as control (CON) were included in this study. Both groups were characterized non-destructively by Stereomicroscopy, Optical Profilometry (Sa, Sq, Sz, Sc), and SEM/EDX analysis and then destructively after metallogaphic preparation employing instrumented indentation testing (HM, EIT, ηIT, and HV) and SEM/EDX at bone-implant interface. Results All retrieved implants showed a loss of gloss with the formation of bone-like formation on the majority of them. However, no differences in surface roughness parameters were identified between macroscopically intact and retrieved regions of implants. The elements precipitated on the surface were O, C, Ca, and P while traces of Na, K, Al, S, Cl, and Mg were also identified. The surface of control sample is characterized by small pits while only Ti and Al traces were identified by EDX analysis. The presence of all the aforementioned elements apart from Ti and Al on the retrieved implants' surface should be appended to the contact of implant with bone and biological fluids while Interfacial analysis revealed a well-formed bone-implant interface. However, no significant differences were found for all mechanical properties tested between RET and CON groups. Conclusions The results of this study indicate that retrieved palatal implant surface has undergone morphological and elemental alterations probably associated with the osseointegration process during service. Insertion and functional loading did not affect the mechanical properties of implants tested.

Statistics

Citations

Dimensions.ai Metrics
2 citations in Web of Science®
2 citations in Scopus®
Google Scholar™

Altmetrics

Downloads

87 downloads since deposited on 13 Feb 2018
19 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > Center for Dental Medicine > Clinic for Orthodontics and Pediatric Dentistry
Dewey Decimal Classification:610 Medicine & health
Scopus Subject Areas:Health Sciences > Orthodontics
Language:English
Date:30 November 2017
Deposited On:13 Feb 2018 13:00
Last Modified:23 Mar 2023 09:23
Publisher:Oxford University Press
ISSN:0141-5387
OA Status:Hybrid
Free access at:Publisher DOI. An embargo period may apply.
Publisher DOI:https://doi.org/10.1093/ejo/cjx001
PubMed ID:28199512
  • Content: Accepted Version