Incidence and Clinical Impact of Device-Associated Thrombus and Peri-Device Leak Following Left Atrial Appendage Closure With the Amplatzer Cardiac Plug


Saw, Jacqueline; Tzikas, Apostolos; Shakir, Samera; Gafoor, Sameer; Omran, Heyder; Nielsen-Kudsk, Jens Erik; Kefer, Joelle; Aminian, Adel; Berti, Sergio; Santoro, Gennaro; Nietlispach, Fabian; Moschovitis, Aris; Cruz-Gonzalez, Ignacio; Stammen, Francis; Tichelbäcker, Tobias; Freixa, Xavier; Ibrahim, Reda; Schillinger, Wolfgang; Meier, Bernhard; Sievert, Horst; Gloekler, Steffen (2017). Incidence and Clinical Impact of Device-Associated Thrombus and Peri-Device Leak Following Left Atrial Appendage Closure With the Amplatzer Cardiac Plug. JACC. Cardiovascular interventions, 10(4):391-399.

Abstract

OBJECTIVES Routine device surveillance after successful left atrial appendage closure is recommended to evaluate for intermediate to late complications. The aim of this study was to assess the incidence and clinical impact of these complications on cardiovascular events. METHODS Centers participating in the Amplatzer Cardiac Plug multicenter study were requested to submit their post-procedural transesophageal echocardiograms for independent adjudication. Thirteen of 22 centers contributed all their post-procedural echocardiograms, which included 344 from 605 consecutive patients. These images were submitted to a core laboratory and reviewed by 2 independent experts for peri-device leak, device-associated thrombus, device embolization, device migration, left atrial appendage thrombus, and left atrial thrombus. Clinical events were prospectively collected by each center. RESULTS Of the 344 transesophageal echocardiograms, 339 were deemed analyzable. Patients' mean age was 74.4 ± 7.5 years, and 67.3% were men. The mean CHADSscore was 2.7 ± 1.3, the mean CHADS-VASc score was 4.3 ± 1.5, and the mean HAS-BLED score was 3.0 ± 1.2. Amplatzer Cardiac Plug implantation was successful in all patients. Periprocedural major adverse events occurred in 2.4%. Median clinical follow-up duration was 355 days (range 179 to 622 days). Follow-up transesophageal echocardiography was performed after a median of 134 days (range 88 to 227 days). Device-associated thrombus was observed in 3.2% and peri-device leak in 12.5% (5.5% minimal, 5.8% mild, 0.6% moderate, 0.6% severe). Neither device-associated thrombus nor peri-device leak was associated with an increased risk for cardiovascular events. Independent predictors of device-associated thrombus were smoking (odds ratio: 5.79; p = 0.017) and female sex (odds ratio: 4.22; p = 0.027). CONCLUSIONS Following successful left atrial appendage closure with the Amplatzer Cardiac Plug, the presence of peri-device leak was relatively low, and device-associated thrombus was infrequent. Neither was associated with increased risk for thromboembolism.

Abstract

OBJECTIVES Routine device surveillance after successful left atrial appendage closure is recommended to evaluate for intermediate to late complications. The aim of this study was to assess the incidence and clinical impact of these complications on cardiovascular events. METHODS Centers participating in the Amplatzer Cardiac Plug multicenter study were requested to submit their post-procedural transesophageal echocardiograms for independent adjudication. Thirteen of 22 centers contributed all their post-procedural echocardiograms, which included 344 from 605 consecutive patients. These images were submitted to a core laboratory and reviewed by 2 independent experts for peri-device leak, device-associated thrombus, device embolization, device migration, left atrial appendage thrombus, and left atrial thrombus. Clinical events were prospectively collected by each center. RESULTS Of the 344 transesophageal echocardiograms, 339 were deemed analyzable. Patients' mean age was 74.4 ± 7.5 years, and 67.3% were men. The mean CHADSscore was 2.7 ± 1.3, the mean CHADS-VASc score was 4.3 ± 1.5, and the mean HAS-BLED score was 3.0 ± 1.2. Amplatzer Cardiac Plug implantation was successful in all patients. Periprocedural major adverse events occurred in 2.4%. Median clinical follow-up duration was 355 days (range 179 to 622 days). Follow-up transesophageal echocardiography was performed after a median of 134 days (range 88 to 227 days). Device-associated thrombus was observed in 3.2% and peri-device leak in 12.5% (5.5% minimal, 5.8% mild, 0.6% moderate, 0.6% severe). Neither device-associated thrombus nor peri-device leak was associated with an increased risk for cardiovascular events. Independent predictors of device-associated thrombus were smoking (odds ratio: 5.79; p = 0.017) and female sex (odds ratio: 4.22; p = 0.027). CONCLUSIONS Following successful left atrial appendage closure with the Amplatzer Cardiac Plug, the presence of peri-device leak was relatively low, and device-associated thrombus was infrequent. Neither was associated with increased risk for thromboembolism.

Statistics

Citations

Dimensions.ai Metrics
17 citations in Web of Science®
23 citations in Scopus®
19 citations in Microsoft Academic
Google Scholar™

Altmetrics

Downloads

8 downloads since deposited on 07 Mar 2018
8 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Clinic for Cardiology
Dewey Decimal Classification:610 Medicine & health
Date:27 February 2017
Deposited On:07 Mar 2018 15:37
Last Modified:14 Mar 2018 18:08
Publisher:Elsevier
ISSN:1876-7605
OA Status:Green
Free access at:Publisher DOI. An embargo period may apply.
Publisher DOI:https://doi.org/10.1016/j.jcin.2016.11.029
PubMed ID:28231907