Precise and Automatic Patient Positioning in Computed Tomography: Avatar Modeling of the Patient Surface Using a 3-Dimensional Camera


Saltybaeva, Natalia; Schmidt, Bernhard; Wimmer, Andreas; Flohr, Thomas; Alkadhi, Hatem (2018). Precise and Automatic Patient Positioning in Computed Tomography: Avatar Modeling of the Patient Surface Using a 3-Dimensional Camera. Investigative Radiology, 53(11):641-646.

Abstract

OBJECTIVES The aim of this study was to evaluate the accuracy of a 3-dimensional (3D) camera algorithm for automatic and individualized patient positioning based on body surface detection and to compare the results of the 3D camera with manual positioning performed by technologists in routinely obtained chest and abdomen computed tomography (CT) examinations. MATERIALS AND METHODS This study included data of 120 patients undergoing clinically indicated chest (n = 68) and abdomen (n = 52) CT. Fifty-two of the patients were scanned with CT using a table height manually selected by technologists; 68 patients were automatically positioned with the 3D camera, which is based on patient-specific body surface and contour detection. The ground truth table height (TGT) was defined as the table height that aligns the axial center of the patient's body region in the CT scanner isocenter. Off-centering was defined as the difference between the ground truth table height (TGT) and the actual table position used in all CT examinations. The t test was performed to determine significant differences in the vertical offset between automatic and manual positioning. The χ test was used to check whether there was a relationship between patient size and the magnitude of off-centering. RESULTS We found a significant improvement in patient centering (offset 5 ± 3 mm) when using the automatic positioning algorithm with the 3D camera compared with manual positioning (offset 19 ± 10 mm) performed by technologists (P < 0.005). Automatic patient positioning based on the 3D camera reduced the average offset in vertical table position from 19 mm to 7 mm for chest and from 18 mm to 4 mm for abdomen CT. The absolute maximal offset was 39 mm and 43 mm for chest and abdomen CT, respectively, when patients were positioned manually, whereas with automatic positioning using the 3D camera the offset never exceeded 15 mm. In chest CT performed with manual patient positioning, we found a significant correlation between vertical offset greater than 20 mm and patient size (body mass index, >26 kg/m, P < 0.001). In contrast, no such relationship was found for abdomen CT (P = 0.38). CONCLUSIONS Automatic individualized patient positioning using a 3D camera allows for accurate patient centering as compared with manual positioning, which improves radiation dose utilization.

Abstract

OBJECTIVES The aim of this study was to evaluate the accuracy of a 3-dimensional (3D) camera algorithm for automatic and individualized patient positioning based on body surface detection and to compare the results of the 3D camera with manual positioning performed by technologists in routinely obtained chest and abdomen computed tomography (CT) examinations. MATERIALS AND METHODS This study included data of 120 patients undergoing clinically indicated chest (n = 68) and abdomen (n = 52) CT. Fifty-two of the patients were scanned with CT using a table height manually selected by technologists; 68 patients were automatically positioned with the 3D camera, which is based on patient-specific body surface and contour detection. The ground truth table height (TGT) was defined as the table height that aligns the axial center of the patient's body region in the CT scanner isocenter. Off-centering was defined as the difference between the ground truth table height (TGT) and the actual table position used in all CT examinations. The t test was performed to determine significant differences in the vertical offset between automatic and manual positioning. The χ test was used to check whether there was a relationship between patient size and the magnitude of off-centering. RESULTS We found a significant improvement in patient centering (offset 5 ± 3 mm) when using the automatic positioning algorithm with the 3D camera compared with manual positioning (offset 19 ± 10 mm) performed by technologists (P < 0.005). Automatic patient positioning based on the 3D camera reduced the average offset in vertical table position from 19 mm to 7 mm for chest and from 18 mm to 4 mm for abdomen CT. The absolute maximal offset was 39 mm and 43 mm for chest and abdomen CT, respectively, when patients were positioned manually, whereas with automatic positioning using the 3D camera the offset never exceeded 15 mm. In chest CT performed with manual patient positioning, we found a significant correlation between vertical offset greater than 20 mm and patient size (body mass index, >26 kg/m, P < 0.001). In contrast, no such relationship was found for abdomen CT (P = 0.38). CONCLUSIONS Automatic individualized patient positioning using a 3D camera allows for accurate patient centering as compared with manual positioning, which improves radiation dose utilization.

Statistics

Citations

Dimensions.ai Metrics
51 citations in Web of Science®
62 citations in Scopus®
Google Scholar™

Altmetrics

Downloads

1382 downloads since deposited on 19 Jun 2018
288 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Clinic for Diagnostic and Interventional Radiology
Dewey Decimal Classification:610 Medicine & health
Scopus Subject Areas:Health Sciences > Radiology, Nuclear Medicine and Imaging
Uncontrolled Keywords:Radiology Nuclear Medicine and imaging, General Medicine
Language:English
Date:1 November 2018
Deposited On:19 Jun 2018 11:23
Last Modified:27 Nov 2023 08:05
Publisher:Lippincott Williams & Wilkins
ISSN:0020-9996
OA Status:Green
Publisher DOI:https://doi.org/10.1097/RLI.0000000000000482
PubMed ID:29762259
  • Content: Published Version
  • Licence: Creative Commons: Attribution-NonCommercial-NoDerivatives 4.0 International (CC BY-NC-ND 4.0)