In vivo precision assessment of a near-infrared spectroscopy-based tissue oximeter (OxyPrem v1.3) in neonates considering systemic hemodynamic fluctuations


Kleiser, Stefan; Ostojic, Daniel; Nasseri, Nassim; Isler, Helene; Bucher, Hans Ulrich; Bassler, Dirk; Wolf, Martin; Scholkmann, Felix; Karen, Tanja (2018). In vivo precision assessment of a near-infrared spectroscopy-based tissue oximeter (OxyPrem v1.3) in neonates considering systemic hemodynamic fluctuations. Journal of Biomedical Optics, 23(06):067003.

Abstract

The aim was to determine the precision of a noninvasive near-infrared spectroscopy (NIRS)-based tissue oximeter (OxyPrem v1.3). Using a linear mixed-effects model, we quantified the variability for cerebral tissue oxygenation (StO2) measurements in 35 preterm neonates to be 2.64%, a value that meets the often-articulated clinicians’ demand for a precise tissue oxygenation measurement. We showed that the variability of StO2 values measured was dominated by spontaneous systemic hemodynamic fluctuations during the measurement, meaning that precision of the instrument was actually even better. Based on simultaneous and continuous measurements of peripheral arterial oxygenation and cerebral StO2 with a second sensor, we were able to determine and quantify the physiological instability precisely. We presented different methods and analyses aiming at reducing this systematic physiological error of in vivo precision assessments. Using these methods, we estimated the precision of the OxyPrem tissue oximeter to be   ≤  1.85  %  . With our study, we deliver relevant information to establish highly precise cerebral oxygenation measurements with NIRS-based oximetry, facilitating the further development toward a substantially improved diagnosis and treatment of patients with respect to brain oxygenation.

Abstract

The aim was to determine the precision of a noninvasive near-infrared spectroscopy (NIRS)-based tissue oximeter (OxyPrem v1.3). Using a linear mixed-effects model, we quantified the variability for cerebral tissue oxygenation (StO2) measurements in 35 preterm neonates to be 2.64%, a value that meets the often-articulated clinicians’ demand for a precise tissue oxygenation measurement. We showed that the variability of StO2 values measured was dominated by spontaneous systemic hemodynamic fluctuations during the measurement, meaning that precision of the instrument was actually even better. Based on simultaneous and continuous measurements of peripheral arterial oxygenation and cerebral StO2 with a second sensor, we were able to determine and quantify the physiological instability precisely. We presented different methods and analyses aiming at reducing this systematic physiological error of in vivo precision assessments. Using these methods, we estimated the precision of the OxyPrem tissue oximeter to be   ≤  1.85  %  . With our study, we deliver relevant information to establish highly precise cerebral oxygenation measurements with NIRS-based oximetry, facilitating the further development toward a substantially improved diagnosis and treatment of patients with respect to brain oxygenation.

Statistics

Citations

Dimensions.ai Metrics

Altmetrics

Downloads

86 downloads since deposited on 19 Jun 2018
76 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Clinic for Neonatology
Dewey Decimal Classification:610 Medicine & health
Uncontrolled Keywords:Atomic and Molecular Physics, and Optics, Electronic, Optical and Magnetic Materials, Biomaterials, Biomedical Engineering
Language:English
Date:2018
Deposited On:19 Jun 2018 10:18
Last Modified:08 Sep 2018 03:11
Publisher:SPIE - International Society for Optical Engineering
ISSN:1083-3668
OA Status:Green
Free access at:Publisher DOI. An embargo period may apply.
Publisher DOI:https://doi.org/10.1117/1.JBO.23.6.067003
PubMed ID:29905038