Separation efficiency versus column length. An experimental study with capillary columns


Grob, K; Grob, G (1969). Separation efficiency versus column length. An experimental study with capillary columns. Journal of Chromatographic Science, 7(8):515-516.

Abstract

The columns of different length used for studying the relation between HETP and column length are supposed to exhibit identical characteristics per meter. In the laboratory this condition cannot be rigorously fulfilled. The best approximation probably consists in starting the work with a relatively long capillary column followed by running shorter parts obtained by progressive cutting of the original column. Based on this procedure a marked increase of HETP with increasing column length is observed

Abstract

The columns of different length used for studying the relation between HETP and column length are supposed to exhibit identical characteristics per meter. In the laboratory this condition cannot be rigorously fulfilled. The best approximation probably consists in starting the work with a relatively long capillary column followed by running shorter parts obtained by progressive cutting of the original column. Based on this procedure a marked increase of HETP with increasing column length is observed

Statistics

Citations

Dimensions.ai Metrics
19 citations in Web of Science®
19 citations in Scopus®
Google Scholar™

Altmetrics

Downloads

20 downloads since deposited on 18 Oct 2018
7 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:National licences > 142-005
Dewey Decimal Classification:Unspecified
Scopus Subject Areas:Physical Sciences > Analytical Chemistry
Language:English
Date:1 August 1969
Deposited On:18 Oct 2018 14:11
Last Modified:15 Apr 2021 14:48
Publisher:Preston Publications
ISSN:0021-9665
OA Status:Green
Free access at:Publisher DOI. An embargo period may apply.
Publisher DOI:https://doi.org/10.1093/chromsci/7.8.515