Determination of left ventricular wall thickness and muscle mass by intravenous digital subtractionangiocardiography: validation of the method


Grob, D; Hess, O M; Monrad, E; Birchler, B; Grimm, J; Krayenbuehl, H P (1988). Determination of left ventricular wall thickness and muscle mass by intravenous digital subtractionangiocardiography: validation of the method. European Heart Journal, 9(1):73-86.

Abstract

Left ventricular (LV) wall thickness and muscle mass are important measures of LV hypertrophy. In 24 patients LV end-diastolic wall thickness and muscle mass were determined (two observers) by digital subtraction angiocardiography (DSA) and conventional LV angiocardiography (LVA). Wall thickness was determined over the anterolateral wall of the left ventricle according to the technique of Rackley (method 1) or by planimetry (method 2). Seventeen patients were studied at rest and seven during dynamic exercise. Wall thickness correlated well between LVA and DSA; the best correlations were obtained by a combined subtraction mode using either method 1 or 2 (method 1, r≥0-80; method2, r≥0. 75). The standard error of estimate of the mean (SEE) was slightly lower for method 2 (≤ 10%) than for method 1 (≤ 13%). DSA significantly overestimated wall thickness by 5-7% with method 1 and underestimated by 12-14% with method 2. Muscle mass correlated well between LVA and DSA; the SEE was ≤ 15% for method 1 and≤ 12% for method 2. Overestimation of muscle mass by DSA was 7-11% with method 1 and underestimation was 13-15% with method 2.It is concluded that LV wall thickness can be determined accurately by DSA with an SEE ranging between 10 and 13%. Determination of LV muscle mass is slightly less accurate and the SEE is slightly larger ranging between 13 to 17%. With method 1, wall thickness and muscle mass were over estimated and with method 2 underestimated

Abstract

Left ventricular (LV) wall thickness and muscle mass are important measures of LV hypertrophy. In 24 patients LV end-diastolic wall thickness and muscle mass were determined (two observers) by digital subtraction angiocardiography (DSA) and conventional LV angiocardiography (LVA). Wall thickness was determined over the anterolateral wall of the left ventricle according to the technique of Rackley (method 1) or by planimetry (method 2). Seventeen patients were studied at rest and seven during dynamic exercise. Wall thickness correlated well between LVA and DSA; the best correlations were obtained by a combined subtraction mode using either method 1 or 2 (method 1, r≥0-80; method2, r≥0. 75). The standard error of estimate of the mean (SEE) was slightly lower for method 2 (≤ 10%) than for method 1 (≤ 13%). DSA significantly overestimated wall thickness by 5-7% with method 1 and underestimated by 12-14% with method 2. Muscle mass correlated well between LVA and DSA; the SEE was ≤ 15% for method 1 and≤ 12% for method 2. Overestimation of muscle mass by DSA was 7-11% with method 1 and underestimation was 13-15% with method 2.It is concluded that LV wall thickness can be determined accurately by DSA with an SEE ranging between 10 and 13%. Determination of LV muscle mass is slightly less accurate and the SEE is slightly larger ranging between 13 to 17%. With method 1, wall thickness and muscle mass were over estimated and with method 2 underestimated

Statistics

Citations

Dimensions.ai Metrics
4 citations in Web of Science®
4 citations in Scopus®
Google Scholar™

Altmetrics

Downloads

8 downloads since deposited on 17 Oct 2018
8 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:National licences > 142-005
Dewey Decimal Classification:570 Life sciences; biology
610 Medicine & health
Language:English
Date:1 January 1988
Deposited On:17 Oct 2018 15:24
Last Modified:01 Nov 2018 14:47
Publisher:Oxford University Press
ISSN:0195-668X
OA Status:Green
Free access at:Publisher DOI. An embargo period may apply.
Publisher DOI:https://doi.org/10.1093/oxfordjournals.eurheartj.a062394