MR-guided percutaneous angioplasty: Assessment of tracking safety, catheter handling and functionality


Wildermuth, Simon; Dumoulin, Charles L; Pfammatter, Thomas; Maier, Stephan E; Hofmann, Eugen; Debatin, Jörg F (1998). MR-guided percutaneous angioplasty: Assessment of tracking safety, catheter handling and functionality. Cardiovascular and Interventional Radiology, 21(5):404-410.

Abstract

Purpose: Magnetic resonance (MR)-guided percutaneous vascular interventions have evolved to a practical possibility with the advent of open-configuration MR systems and real-time tracking techniques. The purpose of this study was to assess an MR-tracking percutaneous transluminal angioplasty (PTA) catheter with regard to its safety profile and functionality. Methods: Real-time, biplanar tracking of the PTA catheter was made possible by incorporating a small radiofrequency (RF) coil in the catheter tip and connecting it to a coaxial cable embedded in the catheter wall. To evaluate potentially hazardous thermal effects due to the incorporation of the coil, temperature measurements were performed within and around the coil under various canning and tracking conditions at 1.5 Tesla (T). Catheter force transmission and balloon-burst pressure of the MR-tracking PTA catheter were compared with those of a standard PTA catheter. The dilatative capability of the angioplasty balloon was assessed in vitro as well as in vivo, in an isolated femoral artery segment in a swine. Results: The degree of heating at the RF coil was directly proportional to the power of the RF pulses. Heating was negligible with MR tracking, conventional spin-echo and low-flip gradient-echo sequences. Sequences with higher duty cycles, such as fast spin echo, produced harmful heating effects. Force transmission of the MR-tracking PTA catheter was slightly inferior to that of the standard PTA catheter, while balloon-burst pressures were similar to those of conventional catheters. The MR-tracking PTA catheter functioned well both in vitro and in vivo. Conclusion: The in vivo use of an MR-tracking PTA catheter is safe under most scanning conditions

Abstract

Purpose: Magnetic resonance (MR)-guided percutaneous vascular interventions have evolved to a practical possibility with the advent of open-configuration MR systems and real-time tracking techniques. The purpose of this study was to assess an MR-tracking percutaneous transluminal angioplasty (PTA) catheter with regard to its safety profile and functionality. Methods: Real-time, biplanar tracking of the PTA catheter was made possible by incorporating a small radiofrequency (RF) coil in the catheter tip and connecting it to a coaxial cable embedded in the catheter wall. To evaluate potentially hazardous thermal effects due to the incorporation of the coil, temperature measurements were performed within and around the coil under various canning and tracking conditions at 1.5 Tesla (T). Catheter force transmission and balloon-burst pressure of the MR-tracking PTA catheter were compared with those of a standard PTA catheter. The dilatative capability of the angioplasty balloon was assessed in vitro as well as in vivo, in an isolated femoral artery segment in a swine. Results: The degree of heating at the RF coil was directly proportional to the power of the RF pulses. Heating was negligible with MR tracking, conventional spin-echo and low-flip gradient-echo sequences. Sequences with higher duty cycles, such as fast spin echo, produced harmful heating effects. Force transmission of the MR-tracking PTA catheter was slightly inferior to that of the standard PTA catheter, while balloon-burst pressures were similar to those of conventional catheters. The MR-tracking PTA catheter functioned well both in vitro and in vivo. Conclusion: The in vivo use of an MR-tracking PTA catheter is safe under most scanning conditions

Statistics

Citations

Dimensions.ai Metrics
63 citations in Web of Science®
72 citations in Scopus®
Google Scholar™

Altmetrics

Downloads

20 downloads since deposited on 26 Sep 2018
20 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:National licences > 142-005
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:1 September 1998
Deposited On:26 Sep 2018 12:50
Last Modified:26 Apr 2019 17:49
Publisher:Springer
ISSN:0174-1551
OA Status:Green
Free access at:Publisher DOI. An embargo period may apply.
Publisher DOI:https://doi.org/10.1007/s002709900288
Related URLs:https://www.swissbib.ch/Search/Results?lookfor=nationallicencespringer101007s002709900288 (Library Catalogue)

Download

Download PDF  'MR-guided percutaneous angioplasty: Assessment of tracking safety, catheter handling and functionality'.
Preview
Content: Published Version
Language: English
Filetype: PDF (Nationallizenz 142-005)
Size: 1MB
View at publisher