Fluid leakage past tracheal tube cuffs: evaluation of the new Microcuff endotracheal tube


Dullenkopf, Alexander; Gerber, Andreas; Weiss, Markus (2003). Fluid leakage past tracheal tube cuffs: evaluation of the new Microcuff endotracheal tube. Intensive Care Medicine, 29(10):1849-1853.

Abstract

Objective: This study compared the recently introduced Microcuff endotracheal tube HVLP ICU featuring an ultrathin (7-µm) polyurethane cuff membrane with endotracheal tubes from different manufacturers regarding fluid leakage past the tube cuff. Design: In vitro setup. Measurements and results: The following endotracheal tubes (ID 7.5mm) were compared: Mallinckrodt HiLo, Microcuff HVLP ICU, Portex Profile Soft Seal, Rüsch Super Safety Clear, and Sheridan CF. A vertical PVC trachea model (ID 20mm) was intubated, and cuffs were inflated to 10, 15, 20, 25, 30, and 60cmH2O. Colored water (5ml) was added to the top of the cuff. The amount of leaked fluid past the tube cuff within 5, 10, and 60min was recorded. Experiments were performed four times using two examples of each tube brand. Fluid leakage past tube cuffs occurred in all conventional endotracheal tubes at cuff pressures from 10 to 60cmH2O. In the Microcuff tube cuff pressure fluid leakage was observed within 10min only at 10cmH2O. Results with the Microcuff tube were significantly better than all other tube brands at cuff pressures of 10-30cmH2O. Conclusions: Within the acceptable upper limit for tracheal cuff pressure (25-30cmH2O) the Microcuff endotracheal tube was the only one of the tested tubes to prevent fluid leakage in our in vitro setup. In vivo studies are required to confirm these findings

Abstract

Objective: This study compared the recently introduced Microcuff endotracheal tube HVLP ICU featuring an ultrathin (7-µm) polyurethane cuff membrane with endotracheal tubes from different manufacturers regarding fluid leakage past the tube cuff. Design: In vitro setup. Measurements and results: The following endotracheal tubes (ID 7.5mm) were compared: Mallinckrodt HiLo, Microcuff HVLP ICU, Portex Profile Soft Seal, Rüsch Super Safety Clear, and Sheridan CF. A vertical PVC trachea model (ID 20mm) was intubated, and cuffs were inflated to 10, 15, 20, 25, 30, and 60cmH2O. Colored water (5ml) was added to the top of the cuff. The amount of leaked fluid past the tube cuff within 5, 10, and 60min was recorded. Experiments were performed four times using two examples of each tube brand. Fluid leakage past tube cuffs occurred in all conventional endotracheal tubes at cuff pressures from 10 to 60cmH2O. In the Microcuff tube cuff pressure fluid leakage was observed within 10min only at 10cmH2O. Results with the Microcuff tube were significantly better than all other tube brands at cuff pressures of 10-30cmH2O. Conclusions: Within the acceptable upper limit for tracheal cuff pressure (25-30cmH2O) the Microcuff endotracheal tube was the only one of the tested tubes to prevent fluid leakage in our in vitro setup. In vivo studies are required to confirm these findings

Statistics

Citations

Dimensions.ai Metrics
139 citations in Web of Science®
162 citations in Scopus®
Google Scholar™

Altmetrics

Downloads

37 downloads since deposited on 11 Oct 2018
21 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:National licences > 142-005
Dewey Decimal Classification:610 Medicine & health
Scopus Subject Areas:Health Sciences > Critical Care and Intensive Care Medicine
Language:English
Date:1 October 2003
Deposited On:11 Oct 2018 10:46
Last Modified:15 Apr 2021 14:52
Publisher:Springer
ISSN:0342-4642
OA Status:Green
Free access at:Publisher DOI. An embargo period may apply.
Publisher DOI:https://doi.org/10.1007/s00134-003-1933-6

Download

Green Open Access

Download PDF  'Fluid leakage past tracheal tube cuffs: evaluation of the new Microcuff endotracheal tube'.
Preview
Content: Published Version
Language: English
Filetype: PDF (Nationallizenz 142-005)
Size: 156kB
View at publisher