Safety of intrauterine devices in MRI


Bussmann, Simon; Luechinger, Roger; Froehlich, Johannes M; von Weymarn, Constantin; Reischauer, Carolin; Koh, Dow Mu; Gutzeit, Andreas (2018). Safety of intrauterine devices in MRI. PLoS ONE, 13(10):e0204220.

Abstract

OBJECTIVES

The paucity of safety information on intrauterine devices (IUD) for magnetic resonance imaging (MRI) examinations is clinically relevant. The aim of this study is to evaluate the MRI safety of clinically used IUDs composed of copper/gold and stainless steel at 1.5T and 3.0T.

MATERIALS AND METHODS

We assessed and compared the displacement force, torque effects, presence of imaging artifacts and heating of IUDs composed of copper/gold (western IUDs) and stainless steel (China) on 1.5 and 3.0T MRI systems.

RESULTS

Gold/Copper IUDs can show small deflection angles of 7° ± 7° in the worst-case field gradient of 40T/m (equivalent to magnetic force of 0.5 mN), while the stainless steel IUD experienced significant magnetic force and deflection (Force > 7.5 N; deflection angle 90° ± 1°). Manual rotation and suspension method show no torque effects on gold/copper IUDs but high torque effects were observed by manual rotation on the stainless steel IUD. Heating measurements showed a temperature increase (rescaled to a wbSAR of 4 W/kg) of 1.4°C at 1.5T / 3.4°C at 3.0 T (stainless steel IUD), 3.2°C at 1.5 T / 3.8°C at 3 T (copper/gold IUD), 3.3°C at 1.5 T / 4.8°C at 3 T (copper 1), 3.8°C at 1.5 T / 4.8°C at 3 T (copper 2). The visible imaging artifacts of the copper and gold IUDs at 3 T MRI reach a diameter of 4 mm ± 1 mm, while the stainless steel IUD resulted in artifacts measuring 200mm ± 10 mm when using gradient echo pulse sequences.

CONCLUSIONS

Standard IUDs (copper/gold) can be considered as conditional for MR safety at 1.5 T and 3.0 T, demonstrating at wbSAR up to 4W/kg and a magnetic field gradient of up to 40T/m with minimal imaging artifacts. The stainless steel IUD, however, induces unacceptable artifacts and is potentially harmful to patients during MRI due to high magnetic dislocation forces and torque (MR unsafe).

Abstract

OBJECTIVES

The paucity of safety information on intrauterine devices (IUD) for magnetic resonance imaging (MRI) examinations is clinically relevant. The aim of this study is to evaluate the MRI safety of clinically used IUDs composed of copper/gold and stainless steel at 1.5T and 3.0T.

MATERIALS AND METHODS

We assessed and compared the displacement force, torque effects, presence of imaging artifacts and heating of IUDs composed of copper/gold (western IUDs) and stainless steel (China) on 1.5 and 3.0T MRI systems.

RESULTS

Gold/Copper IUDs can show small deflection angles of 7° ± 7° in the worst-case field gradient of 40T/m (equivalent to magnetic force of 0.5 mN), while the stainless steel IUD experienced significant magnetic force and deflection (Force > 7.5 N; deflection angle 90° ± 1°). Manual rotation and suspension method show no torque effects on gold/copper IUDs but high torque effects were observed by manual rotation on the stainless steel IUD. Heating measurements showed a temperature increase (rescaled to a wbSAR of 4 W/kg) of 1.4°C at 1.5T / 3.4°C at 3.0 T (stainless steel IUD), 3.2°C at 1.5 T / 3.8°C at 3 T (copper/gold IUD), 3.3°C at 1.5 T / 4.8°C at 3 T (copper 1), 3.8°C at 1.5 T / 4.8°C at 3 T (copper 2). The visible imaging artifacts of the copper and gold IUDs at 3 T MRI reach a diameter of 4 mm ± 1 mm, while the stainless steel IUD resulted in artifacts measuring 200mm ± 10 mm when using gradient echo pulse sequences.

CONCLUSIONS

Standard IUDs (copper/gold) can be considered as conditional for MR safety at 1.5 T and 3.0 T, demonstrating at wbSAR up to 4W/kg and a magnetic field gradient of up to 40T/m with minimal imaging artifacts. The stainless steel IUD, however, induces unacceptable artifacts and is potentially harmful to patients during MRI due to high magnetic dislocation forces and torque (MR unsafe).

Statistics

Citations

Altmetrics

Downloads

8 downloads since deposited on 04 Dec 2018
8 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > Institute of Biomedical Engineering
Dewey Decimal Classification:170 Ethics
610 Medicine & health
Language:English
Date:2018
Deposited On:04 Dec 2018 19:41
Last Modified:04 Dec 2018 19:41
Publisher:Public Library of Science (PLoS)
ISSN:1932-6203
OA Status:Gold
Free access at:Publisher DOI. An embargo period may apply.
Publisher DOI:https://doi.org/10.1371/journal.pone.0204220
PubMed ID:30300364

Download

Download PDF  'Safety of intrauterine devices in MRI'.
Preview
Content: Published Version
Language: English
Filetype: PDF
Size: 15MB
View at publisher
Licence: Creative Commons: Attribution 4.0 International (CC BY 4.0)