Interdevice variability of central corneal thickness measurement


Abstract

PURPOSE To evaluate variability of central corneal thickness measurement (CCT) devices using a hitherto unprecedented number of CCT devices.
METHODS CCT was measured consecutively in 122 normal corneas of 61 subjects with seven different devices using three distinct measurement technologies: Scheimpflug, Ultrasound, and Optical Coherence Tomography (OCT). Per device deviation from the mean CCT value per eye was used to determine which of the devices performed best, compared to the mean value.
RESULTS Cirrus OCT yielded the lowest deviation. Deviations of the individual devices from the mean CCT of each eye were (OS/OD) 12.8±5.0/14.9±9.4 μm for Topcon noncontact specular microscopy (NCSM), 11.3±5.9/10.6±7.3 μm for Pentacam, 10.7±5.2/10.4±4.8 μm for Spectralis OCT, 6.0±3.9/6.2±4.9 μm for Topcon DRI OCT, 5.1±3.4/5.9±10.3 μm for AngioVue OCT, 4.8±4.1/5.7±4.6 μm for US pachymetry, and 4.2±3.2/5.7±4.6 μm for Cirrus OCT. The maximum differences between US pachymetry and the other devices were very high (up to 120 μm).
CONCLUSION Central corneal thickness may be under- or overestimated due to high interdevice variations. Measuring CCT with one device only may lead to inappropriate clinical and surgical recommendations. OCT showed superior results.

Abstract

PURPOSE To evaluate variability of central corneal thickness measurement (CCT) devices using a hitherto unprecedented number of CCT devices.
METHODS CCT was measured consecutively in 122 normal corneas of 61 subjects with seven different devices using three distinct measurement technologies: Scheimpflug, Ultrasound, and Optical Coherence Tomography (OCT). Per device deviation from the mean CCT value per eye was used to determine which of the devices performed best, compared to the mean value.
RESULTS Cirrus OCT yielded the lowest deviation. Deviations of the individual devices from the mean CCT of each eye were (OS/OD) 12.8±5.0/14.9±9.4 μm for Topcon noncontact specular microscopy (NCSM), 11.3±5.9/10.6±7.3 μm for Pentacam, 10.7±5.2/10.4±4.8 μm for Spectralis OCT, 6.0±3.9/6.2±4.9 μm for Topcon DRI OCT, 5.1±3.4/5.9±10.3 μm for AngioVue OCT, 4.8±4.1/5.7±4.6 μm for US pachymetry, and 4.2±3.2/5.7±4.6 μm for Cirrus OCT. The maximum differences between US pachymetry and the other devices were very high (up to 120 μm).
CONCLUSION Central corneal thickness may be under- or overestimated due to high interdevice variations. Measuring CCT with one device only may lead to inappropriate clinical and surgical recommendations. OCT showed superior results.

Statistics

Citations

Dimensions.ai Metrics
12 citations in Web of Science®
14 citations in Scopus®
Google Scholar™

Altmetrics

Downloads

9 downloads since deposited on 11 Jan 2019
2 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Ophthalmology Clinic
Dewey Decimal Classification:610 Medicine & health
Scopus Subject Areas:Life Sciences > General Biochemistry, Genetics and Molecular Biology
Life Sciences > General Agricultural and Biological Sciences
Health Sciences > Multidisciplinary
Language:English
Date:2018
Deposited On:11 Jan 2019 13:23
Last Modified:01 Dec 2023 08:03
Publisher:Public Library of Science (PLoS)
ISSN:1932-6203
OA Status:Gold
Free access at:PubMed ID. An embargo period may apply.
Publisher DOI:https://doi.org/10.1371/journal.pone.0203884
PubMed ID:30212550
  • Content: Published Version
  • Licence: Creative Commons: Attribution 4.0 International (CC BY 4.0)