Surgical Atrioventricular Valve Replacement With Melody Valve in Infants and Children


Abstract

Background Pediatric patients with atrioventricular valve disease have limited options for prosthetic valve replacement in sizes <15 mm. Based on successful experience with the stented bovine jugular vein graft (Melody valve) in the right ventricular outflow tract, the prosthesis has been modified for surgical valve replacement in pediatric patients with atrioventricular dysfunction with the intention of subsequent valve expansion in the catheterization laboratory as the child grows. Methods and Results A multicenter, retrospective cohort study was performed among patients who underwent atrioventricular valve replacement with Melody valve at 17 participating sites from North America and Europe, including 68 patients with either mitral (n=59) or tricuspid (n=9) replacement at a median age of 8 months (range, 3 days to 13 years). The median size at implantation was 14 mm (range, 9-24 mm). Immediately postoperatively, the valve was competent with low gradients in all patients. Fifteen patients died; 3 patients underwent transplantation. Nineteen patients required reoperation for adverse outcomes, including valve explantation (n=16), left ventricular outflow tract obstruction (n=1), permanent pacemaker implantation (n=1), and paravalvular leak repair (n=1). Twenty-five patients underwent 41 episodes of catheter-based balloon expansion, exhibiting a significant decrease in median gradient ( P<0.001) with no significant increase in grade of regurgitation. Twelve months after implantation, cumulative incidence analysis indicated that 55% of the patients would be expected to be free from death, heart transplantation, structural valve deterioration, or valve replacement. Conclusions The Melody valve is a feasible option for surgical atrioventricular valve replacement in patients with hypoplastic annuli. The prosthesis shows acceptable short-term function and is amenable to catheter-based enlargement as the child grows. However, patients remain at risk for mortality and structural valve deterioration, despite adequate early valvular function. Device design and implantation techniques must be refined to reduce complications and extend durability. Clinical Trial Registration URL: https://www.clinicaltrials.gov. Unique identifier: NCT02505074.

Abstract

Background Pediatric patients with atrioventricular valve disease have limited options for prosthetic valve replacement in sizes <15 mm. Based on successful experience with the stented bovine jugular vein graft (Melody valve) in the right ventricular outflow tract, the prosthesis has been modified for surgical valve replacement in pediatric patients with atrioventricular dysfunction with the intention of subsequent valve expansion in the catheterization laboratory as the child grows. Methods and Results A multicenter, retrospective cohort study was performed among patients who underwent atrioventricular valve replacement with Melody valve at 17 participating sites from North America and Europe, including 68 patients with either mitral (n=59) or tricuspid (n=9) replacement at a median age of 8 months (range, 3 days to 13 years). The median size at implantation was 14 mm (range, 9-24 mm). Immediately postoperatively, the valve was competent with low gradients in all patients. Fifteen patients died; 3 patients underwent transplantation. Nineteen patients required reoperation for adverse outcomes, including valve explantation (n=16), left ventricular outflow tract obstruction (n=1), permanent pacemaker implantation (n=1), and paravalvular leak repair (n=1). Twenty-five patients underwent 41 episodes of catheter-based balloon expansion, exhibiting a significant decrease in median gradient ( P<0.001) with no significant increase in grade of regurgitation. Twelve months after implantation, cumulative incidence analysis indicated that 55% of the patients would be expected to be free from death, heart transplantation, structural valve deterioration, or valve replacement. Conclusions The Melody valve is a feasible option for surgical atrioventricular valve replacement in patients with hypoplastic annuli. The prosthesis shows acceptable short-term function and is amenable to catheter-based enlargement as the child grows. However, patients remain at risk for mortality and structural valve deterioration, despite adequate early valvular function. Device design and implantation techniques must be refined to reduce complications and extend durability. Clinical Trial Registration URL: https://www.clinicaltrials.gov. Unique identifier: NCT02505074.

Statistics

Citations

Dimensions.ai Metrics
2 citations in Web of Science®
2 citations in Scopus®
Google Scholar™

Altmetrics

Downloads

0 downloads since deposited on 05 Feb 2019
0 downloads since 12 months

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Children's Hospital Zurich > Medical Clinic
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:November 2018
Deposited On:05 Feb 2019 20:36
Last Modified:05 Feb 2019 20:52
Publisher:Lippincott Williams & Wilkins
ISSN:1941-7640
OA Status:Closed
Free access at:Publisher DOI. An embargo period may apply.
Publisher DOI:https://doi.org/10.1161/CIRCINTERVENTIONS.118.007145
PubMed ID:30571200

Download

Content: Published Version
Filetype: PDF - Registered users only until 16 November 2019
Size: 350kB
View at publisher