Dosimetric influence of pitch in patient positioning for radiotherapy of long treatment volumes; the usefulness of six degree of freedom couch


Stieb, Sonja; Malla, Michelle; Graydon, Shaun; Riesterer, Oliver; Klöck, Stephan; Studer, Gabriela; Tanadini-Lang, Stephanie (2018). Dosimetric influence of pitch in patient positioning for radiotherapy of long treatment volumes; the usefulness of six degree of freedom couch. British Journal of Radiology, 91(1091):20170704.

Abstract

OBJECTIVE:
Pitch, the rotation around the transversal axis of the patient during radiotherapy has little impact on the dose distribution of small spherical treatment volumes; however it might affect treatment of long volumes requiring a correction with a six degree of freedom couch.
METHODS:
We included 10 patients each with nasopharyngeal carcinoma (NPC) and esophageal cancer, treated with volumetric modulated arc therapy. Pitch was simulated by tilting the planning CT in ventral and dorsal direction by ± 1.5° and ± 3°. Verification plans were calculated on the tilted datasets and were compared to the original plan and the dose constraints of the organs at risk (OAR).
RESULTS:
The deviation in dose to the planning target volume is increasing with the degree of pitch with mean changes of up to 2% for NPC and 1% for esophageal cancer. The most affected OAR in NPC patients are brainstem (max. dose +6.0%) and spinal cord (max. dose +10.0%) when tilted by 3° dorsally and lenses (max. dose +3.3%), oral mucosa (mean dose +2.6%) and parotid glands (mean dose +4.3%) when tilted by 3° ventrally. For esophageal cancer patients, there was no significant change in dose to any OAR. Whereas for esophageal cancer, all tilted treatment plans were still clinically acceptable regarding OAR, 5 NPC plans would no longer be acceptable with a pitch of 1.5° ventral (N = 1), 3° ventral (N = 2) and 3° dorsal (N = 2).
CONCLUSION:
Planning target volume coverage in both tumor entities was only slightly affected, but pitch errors could be relevant for OAR in NPC patients.
ADVANCES IN KNOWLEDGE:
A correction with a six degree of freedom couch is recommended for NPC patients with a pitch mismatch of more than 1.5° to avoid exceeded doses to the OAR.

Abstract

OBJECTIVE:
Pitch, the rotation around the transversal axis of the patient during radiotherapy has little impact on the dose distribution of small spherical treatment volumes; however it might affect treatment of long volumes requiring a correction with a six degree of freedom couch.
METHODS:
We included 10 patients each with nasopharyngeal carcinoma (NPC) and esophageal cancer, treated with volumetric modulated arc therapy. Pitch was simulated by tilting the planning CT in ventral and dorsal direction by ± 1.5° and ± 3°. Verification plans were calculated on the tilted datasets and were compared to the original plan and the dose constraints of the organs at risk (OAR).
RESULTS:
The deviation in dose to the planning target volume is increasing with the degree of pitch with mean changes of up to 2% for NPC and 1% for esophageal cancer. The most affected OAR in NPC patients are brainstem (max. dose +6.0%) and spinal cord (max. dose +10.0%) when tilted by 3° dorsally and lenses (max. dose +3.3%), oral mucosa (mean dose +2.6%) and parotid glands (mean dose +4.3%) when tilted by 3° ventrally. For esophageal cancer patients, there was no significant change in dose to any OAR. Whereas for esophageal cancer, all tilted treatment plans were still clinically acceptable regarding OAR, 5 NPC plans would no longer be acceptable with a pitch of 1.5° ventral (N = 1), 3° ventral (N = 2) and 3° dorsal (N = 2).
CONCLUSION:
Planning target volume coverage in both tumor entities was only slightly affected, but pitch errors could be relevant for OAR in NPC patients.
ADVANCES IN KNOWLEDGE:
A correction with a six degree of freedom couch is recommended for NPC patients with a pitch mismatch of more than 1.5° to avoid exceeded doses to the OAR.

Statistics

Citations

Dimensions.ai Metrics
3 citations in Web of Science®
3 citations in Scopus®
Google Scholar™

Altmetrics

Downloads

22 downloads since deposited on 25 Jan 2019
16 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Clinic for Radiation Oncology
Dewey Decimal Classification:610 Medicine & health
Scopus Subject Areas:Health Sciences > Radiology, Nuclear Medicine and Imaging
Language:English
Date:November 2018
Deposited On:25 Jan 2019 11:36
Last Modified:29 Jul 2020 09:23
Publisher:British Institute of Radiology
ISSN:0007-1285
OA Status:Green
Free access at:PubMed ID. An embargo period may apply.
Publisher DOI:https://doi.org/10.1259/bjr.20170704
PubMed ID:30004794

Download

Green Open Access

Download PDF  'Dosimetric influence of pitch in patient positioning for radiotherapy of long treatment volumes; the usefulness of six degree of freedom couch'.
Preview
Content: Published Version
Filetype: PDF
Size: 990kB
View at publisher