Filling the dead-time gap in zero echo time MRI: Principles compared


Froidevaux, Romain; Weiger, Markus; Brunner, David O; Dietrich, Benjamin E; Wilm, Bertram J; Pruessmann, Klaas P (2018). Filling the dead-time gap in zero echo time MRI: Principles compared. Magnetic Resonance in Medicine, 79(4):2036-2045.

Abstract

PURPOSE:
MRI of tissues with short coherence lifetimes T2 or T2* can be performed efficiently using zero echo time (ZTE) techniques such as algebraic ZTE, pointwise encoding time reduction with radial acquisition (PETRA), and water- and fat-suppressed proton projection MRI (WASPI). They share the principal challenge of recovering data in central k-space missed due to an initial radiofrequency dead time. The purpose of this study was to compare the three techniques directly, with a particular focus on their behavior in the presence of ultra-short-lived spins.
METHODS:
The most direct comparison was enabled by aligning acquisition and reconstruction strategies of the three techniques. Image quality and short- T2* performance were investigated using point spread functions, 3D simulations, and imaging of phantom and bone samples with short (<1 ms) and ultra-short (<100 μs) T2*.
RESULTS:
Algebraic ZTE offers favorable properties but is limited to k-space gaps up to approximately three Nyquist dwells. At larger gaps, PETRA enables robust imaging with little compromise in image quality, whereas WASPI may be prone to artifacts from ultra-short T2* species.
CONCLUSION:
For small k-space gaps (<4 dwells) and T2* much larger than the dead time, all techniques enable artifact-free short- T2* MRI. However, if these requirements are not fulfilled careful consideration is needed and PETRA will generally achieve better image quality. Magn Reson Med 79:2036-2045, 2018. © 2017 International Society for Magnetic Resonance in Medicine.

Abstract

PURPOSE:
MRI of tissues with short coherence lifetimes T2 or T2* can be performed efficiently using zero echo time (ZTE) techniques such as algebraic ZTE, pointwise encoding time reduction with radial acquisition (PETRA), and water- and fat-suppressed proton projection MRI (WASPI). They share the principal challenge of recovering data in central k-space missed due to an initial radiofrequency dead time. The purpose of this study was to compare the three techniques directly, with a particular focus on their behavior in the presence of ultra-short-lived spins.
METHODS:
The most direct comparison was enabled by aligning acquisition and reconstruction strategies of the three techniques. Image quality and short- T2* performance were investigated using point spread functions, 3D simulations, and imaging of phantom and bone samples with short (<1 ms) and ultra-short (<100 μs) T2*.
RESULTS:
Algebraic ZTE offers favorable properties but is limited to k-space gaps up to approximately three Nyquist dwells. At larger gaps, PETRA enables robust imaging with little compromise in image quality, whereas WASPI may be prone to artifacts from ultra-short T2* species.
CONCLUSION:
For small k-space gaps (<4 dwells) and T2* much larger than the dead time, all techniques enable artifact-free short- T2* MRI. However, if these requirements are not fulfilled careful consideration is needed and PETRA will generally achieve better image quality. Magn Reson Med 79:2036-2045, 2018. © 2017 International Society for Magnetic Resonance in Medicine.

Statistics

Citations

Dimensions.ai Metrics
3 citations in Web of Science®
3 citations in Scopus®
Google Scholar™

Altmetrics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > Institute of Biomedical Engineering
Dewey Decimal Classification:170 Ethics
610 Medicine & health
Uncontrolled Keywords:Radiology Nuclear Medicine and imaging
Language:English
Date:1 April 2018
Deposited On:06 Mar 2019 13:11
Last Modified:06 Mar 2019 13:11
Publisher:Wiley-Blackwell Publishing, Inc.
ISSN:0740-3194
OA Status:Closed
Publisher DOI:https://doi.org/10.1002/mrm.26875
PubMed ID:28856717

Download

Full text not available from this repository.
View at publisher

Get full-text in a library