PURPOSE
To test three potential prosthetic material options for zirconia implants in regard to their mechanical properties, loading and retention capacity as well as to record abrasion after chewing simulation followed by thermocyclic aging.
METHODS
Molar crowns (n = 96) of three different computer-aided design/computer-aided manufacturing (CAD/CAM) materials were produced and cemented on zirconia implants (ceramic.implant, Vita) with a diameter of 4.5 mm. Monolithic zirconia (Vita YZ [YZ] with RelyX Unicem 2 Automix [RUN], polymer-infiltrated ceramic (Vita Enamic [VE]) with Vita Adiva F-Cem [VAF] and acrylate polymer (CAD Temp [CT]) with RelyX Ultimate [RUL]. Fracture load and retentive force of the crowns were measured after 24 h water storage at 37 °C and after a chewing simulation followed by thermocyclic aging. Abrasion was recorded by matching stereolithography-data of the crowns obtained before and after chewing simulation. Additionally, the mechanical properties and bonding capabilities of the crown and cement materials were assessed.
RESULTS
Fracture load values were significantly highest for YZ > VE = CT. Retention force values did not differ significantly between the materials. The aging procedure did not affect the fracture load values nor the retention force significantly. Abrasion depth of the crowns was lowest for YZ followed by VE and CT. On unpolished crowns, abrasion of YZ and VE tended to be higher than on polished specimens.
CONCLUSIONS
Based on the obtained in-vitro results, all tested materials can be recommended for the use on zirconia implants, although CT is only approved for temporary crowns. The loading and retention capacity of the materials were not significantly affected by aging.

Abstract

PURPOSE
To test three potential prosthetic material options for zirconia implants in regard to their mechanical properties, loading and retention capacity as well as to record abrasion after chewing simulation followed by thermocyclic aging.
METHODS
Molar crowns (n = 96) of three different computer-aided design/computer-aided manufacturing (CAD/CAM) materials were produced and cemented on zirconia implants (ceramic.implant, Vita) with a diameter of 4.5 mm. Monolithic zirconia (Vita YZ [YZ] with RelyX Unicem 2 Automix [RUN], polymer-infiltrated ceramic (Vita Enamic [VE]) with Vita Adiva F-Cem [VAF] and acrylate polymer (CAD Temp [CT]) with RelyX Ultimate [RUL]. Fracture load and retentive force of the crowns were measured after 24 h water storage at 37 °C and after a chewing simulation followed by thermocyclic aging. Abrasion was recorded by matching stereolithography-data of the crowns obtained before and after chewing simulation. Additionally, the mechanical properties and bonding capabilities of the crown and cement materials were assessed.
RESULTS
Fracture load values were significantly highest for YZ > VE = CT. Retention force values did not differ significantly between the materials. The aging procedure did not affect the fracture load values nor the retention force significantly. Abrasion depth of the crowns was lowest for YZ followed by VE and CT. On unpolished crowns, abrasion of YZ and VE tended to be higher than on polished specimens.
CONCLUSIONS
Based on the obtained in-vitro results, all tested materials can be recommended for the use on zirconia implants, although CT is only approved for temporary crowns. The loading and retention capacity of the materials were not significantly affected by aging.

Statistics

Citations

Dimensions.ai Metrics

Altmetrics

Downloads

4 downloads since deposited on 25 Apr 2019
4 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > Center for Dental Medicine > Clinic of Reconstructive Dentistry
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:1 July 2019
Deposited On:25 Apr 2019 14:11
Last Modified:08 Mar 2020 01:00
Publisher:Elsevier
ISSN:1883-1958
OA Status:Green
Publisher DOI:https://doi.org/10.1016/j.jpor.2019.02.002
PubMed ID:30853580

Download

Green Open Access

Download PDF  'Chewing simulation of zirconia implant supported restorations'.
Preview
Content: Accepted Version
Filetype: PDF
Size: 6MB
View at publisher
Licence: Creative Commons: Attribution-NonCommercial-NoDerivatives 4.0 International (CC BY-NC-ND 4.0)
Content: Published Version
Filetype: PDF - Registered users only
Size: 849kB