Breast Cancer Assessment With Pulse-Echo Speed of Sound Ultrasound From Intrinsic Tissue Reflections: Proof-of-Concept


Ruby, Lisa; Sanabria, Sergio J; Martini, Katharina; Dedes, Konstantin J; Vorburger, Denise; Oezkan, Ece; Frauenfelder, Thomas; Goksel, Orcun; Rominger, Marga B (2019). Breast Cancer Assessment With Pulse-Echo Speed of Sound Ultrasound From Intrinsic Tissue Reflections: Proof-of-Concept. Investigative Radiology, 54(7):419-427.

Abstract

PURPOSE
The aim of this study was to differentiate malignant and benign solid breast lesions with a novel ultrasound (US) technique, which measures speed of sound (SoS) using standard US transducers and intrinsic tissue reflections and scattering (speckles) as internal reference.
MATERIALS AND METHODS
This prospective, institutional review board-approved, Health Insurance Portability and Accountability Act-compliant prospective comparison study was performed with prior written informed consent from 20 women. Ten women with histological proven breast cancer and 10 with fibroadenoma were measured. A conventional US system with a linear probe was used for SoS-US (SonixTouch; Ultrasonix, Richmond, British Columbia, Canada). Tissue speckle reflections served as a timing reference for the US signals transmitted through the breasts. Relative phase inconsistencies were detected using plane wave measurements from different angular directions, and SoS images with 0.5-mm resolution were generated using a spatial domain reconstruction algorithm. The SoS of tumors were compared with the breast density of a larger cohort of 106 healthy women.
RESULTS
Breast lesions show focal increments ΔSoS (meters per second) with respect to the tissue background. Peak ΔSoS values were evaluated. Breast carcinoma showed significantly higher ΔSoS than fibroadenomas ([INCREMENT]SoS > 41.64 m/s: sensitivity, 90%; specificity, 80%; area under curve, 0.910) and healthy breast tissue of different densities (area under curve, 0.938; sensitivity, 90%; specificity, 96.5%). The lesion localization in SoS-US images was consistent with B-mode imaging and repeated SoS-US measurements were reproducible.
CONCLUSIONS
Using SoS-US, based on conventional US and tissue speckles as timing reference, breast carcinoma showed significantly higher SoS values than fibroadenoma and healthy breast tissue of different densities. The SoS presents a promising technique for differentiating solid breast lesions.

Abstract

PURPOSE
The aim of this study was to differentiate malignant and benign solid breast lesions with a novel ultrasound (US) technique, which measures speed of sound (SoS) using standard US transducers and intrinsic tissue reflections and scattering (speckles) as internal reference.
MATERIALS AND METHODS
This prospective, institutional review board-approved, Health Insurance Portability and Accountability Act-compliant prospective comparison study was performed with prior written informed consent from 20 women. Ten women with histological proven breast cancer and 10 with fibroadenoma were measured. A conventional US system with a linear probe was used for SoS-US (SonixTouch; Ultrasonix, Richmond, British Columbia, Canada). Tissue speckle reflections served as a timing reference for the US signals transmitted through the breasts. Relative phase inconsistencies were detected using plane wave measurements from different angular directions, and SoS images with 0.5-mm resolution were generated using a spatial domain reconstruction algorithm. The SoS of tumors were compared with the breast density of a larger cohort of 106 healthy women.
RESULTS
Breast lesions show focal increments ΔSoS (meters per second) with respect to the tissue background. Peak ΔSoS values were evaluated. Breast carcinoma showed significantly higher ΔSoS than fibroadenomas ([INCREMENT]SoS > 41.64 m/s: sensitivity, 90%; specificity, 80%; area under curve, 0.910) and healthy breast tissue of different densities (area under curve, 0.938; sensitivity, 90%; specificity, 96.5%). The lesion localization in SoS-US images was consistent with B-mode imaging and repeated SoS-US measurements were reproducible.
CONCLUSIONS
Using SoS-US, based on conventional US and tissue speckles as timing reference, breast carcinoma showed significantly higher SoS values than fibroadenoma and healthy breast tissue of different densities. The SoS presents a promising technique for differentiating solid breast lesions.

Statistics

Citations

Altmetrics

Downloads

0 downloads since deposited on 02 May 2019
0 downloads since 12 months

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Clinic for Diagnostic and Interventional Radiology
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:July 2019
Deposited On:02 May 2019 10:11
Last Modified:25 Sep 2019 00:33
Publisher:Lippincott Williams & Wilkins
ISSN:0020-9996
OA Status:Closed
Publisher DOI:https://doi.org/10.1097/RLI.0000000000000553
PubMed ID:30913054

Download