Hu, Lien-Hsin; Sharir, Tali; Miller, Robert J H; Einstein, Andrew J; Fish, Mathews B; Ruddy, Terrence D; Dorbala, Sharmila; Di Carli, Marcelo; Kaufmann, Philipp A; Sinusas, Albert J; Miller, Edward J; Bateman, Timothy M; Betancur, Julian; Germano, Guido; Liang, Joanna X; Commandeur, Frederic; Azadani, Peyman N; Gransar, Heidi; Otaki, Yuka; Tamarappoo, Balaji K; Dey, Damini; Berman, Daniel S; Slomka, Piotr J (2019). Upper reference limits of transient ischemic dilation ratio for different protocols on new-generation cadmium zinc telluride cameras: A report from REFINE SPECT registry. Journal of Nuclear Cardiology:Epub ahead of print.

Abstract

BACKGROUND

Upper reference limits for transient ischemic dilation (TID) have not been rigorously established for cadmium-zinc-telluride (CZT) camera systems. We aimed to derive TID limits for common myocardial perfusion imaging protocols utilizing a large, multicenter registry (REFINE SPECT).

METHODS

One thousand six hundred and seventy-two patients with low likelihood of coronary artery disease with normal perfusion findings were identified. Images were processed with Quantitative Perfusion SPECT software (Cedars-Sinai Medical Center, Los Angeles, CA). Non-attenuation-corrected, camera-, radiotracer-, and stress protocol-specific TID limits in supine position were derived from 97.5th percentile and mean + 2 standard deviations (SD). Reference limits were compared for different solid-state cameras (D-SPECT vs. Discovery), radiotracers (technetium-99m-sestamibi vs. tetrofosmin), different types of stress (exercise vs. four different vasodilator-based protocols), and different vasodilator-based protocols.

RESULTS

TID measurements did not follow Gaussian distribution in six out of eight subgroups. TID limits ranged from 1.18 to 1.52 (97.5th percentile) and 1.18 to 1.39 (mean + 2SD). No difference was noted between D-SPECT and Discovery cameras (P = 0.71) while differences between exercise and vasodilator-based protocols (adenosine, regadenoson, or regadenoson-walk) were noted (all P < 0.05).

CONCLUSIONS

We used a multicenter registry to establish camera-, radiotracer-, and protocol-specific upper reference limits of TID for supine position on CZT camera systems. Reference limits did not differ between D-SPECT and Discovery camera.

Abstract

BACKGROUND

Upper reference limits for transient ischemic dilation (TID) have not been rigorously established for cadmium-zinc-telluride (CZT) camera systems. We aimed to derive TID limits for common myocardial perfusion imaging protocols utilizing a large, multicenter registry (REFINE SPECT).

METHODS

One thousand six hundred and seventy-two patients with low likelihood of coronary artery disease with normal perfusion findings were identified. Images were processed with Quantitative Perfusion SPECT software (Cedars-Sinai Medical Center, Los Angeles, CA). Non-attenuation-corrected, camera-, radiotracer-, and stress protocol-specific TID limits in supine position were derived from 97.5th percentile and mean + 2 standard deviations (SD). Reference limits were compared for different solid-state cameras (D-SPECT vs. Discovery), radiotracers (technetium-99m-sestamibi vs. tetrofosmin), different types of stress (exercise vs. four different vasodilator-based protocols), and different vasodilator-based protocols.

RESULTS

TID measurements did not follow Gaussian distribution in six out of eight subgroups. TID limits ranged from 1.18 to 1.52 (97.5th percentile) and 1.18 to 1.39 (mean + 2SD). No difference was noted between D-SPECT and Discovery cameras (P = 0.71) while differences between exercise and vasodilator-based protocols (adenosine, regadenoson, or regadenoson-walk) were noted (all P < 0.05).

CONCLUSIONS

We used a multicenter registry to establish camera-, radiotracer-, and protocol-specific upper reference limits of TID for supine position on CZT camera systems. Reference limits did not differ between D-SPECT and Discovery camera.

Statistics

Citations

Dimensions.ai Metrics

Altmetrics

Downloads

0 downloads since deposited on 31 Jul 2019
0 downloads since 12 months

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Clinic for Nuclear Medicine
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:13 May 2019
Deposited On:31 Jul 2019 13:26
Last Modified:25 Sep 2019 00:39
Publisher:Springer
ISSN:1071-3581
OA Status:Closed
Publisher DOI:https://doi.org/10.1007/s12350-019-01730-y
PubMed ID:31087268

Download

Closed Access: Download allowed only for UZH members