Which Confounders Have the Largest Impact in Shear Wave Elastography of Muscle and How Can They be Minimized? An Elasticity Phantom, Ex Vivo Porcine Muscle and Volunteer Study Using a Commercially Available System


Ruby, Lisa; Mutschler, Tim; Martini, Katharina; Klingmüller, Volker; Frauenfelder, Thomas; Rominger, Marga B; Sanabria, Sergio J (2019). Which Confounders Have the Largest Impact in Shear Wave Elastography of Muscle and How Can They be Minimized? An Elasticity Phantom, Ex Vivo Porcine Muscle and Volunteer Study Using a Commercially Available System. Ultrasound in Medicine & Biology, 45(10):2591-2611.

Abstract

The goal of the study was to investigate the quantitative impact of region of interest (ROI), software choice, muscle fiber orientation and preload tension on shear wave velocity (SWV). First, SWV was assessed in an isotropic elasticity phantom and ex vivo porcine muscle using a commercially available clinical ultrasound system. Secondly, SWV was acquired in relaxed and stretched calf muscles of healthy volunteers (dorsal extension of the talocrural joint), for both parallel and transverse probe direction to the fibers, as well as for different ROIs and software versions. The effect of intermediate probe-fiber alignments was also analyzed. Finally, the impact of confounding factors on SWV reproducibility was minimized with a second force-controlled volunteer study, in which the calf was isometrically loaded, and fiber orientation and ROI were well-defined. 2046 in vivoSWE images were acquired to analyze SWV reproducibility with different confounder settings. In healthy volunteers, the main variance-contributing factors were in order of importance muscle tension, fiber orientation, horizontal ROI size and insertion depth. Regression analysis showed significantly reduced SWV with increasing insertion depth for each study material. Parallel probe-fiber orientation, muscle stretch and increasing horizontal ROI size led to significantly higher SWV. Based on the results of the study, we provide recommendations to minimize the impact of confounders in musculoskeletal elastography and discuss the main confounding mechanisms and trade-offs between confounding variables. Coefficients of variation can be significantly reduced with a controlled protocol, if the confounders are clinically taken into account.

Abstract

The goal of the study was to investigate the quantitative impact of region of interest (ROI), software choice, muscle fiber orientation and preload tension on shear wave velocity (SWV). First, SWV was assessed in an isotropic elasticity phantom and ex vivo porcine muscle using a commercially available clinical ultrasound system. Secondly, SWV was acquired in relaxed and stretched calf muscles of healthy volunteers (dorsal extension of the talocrural joint), for both parallel and transverse probe direction to the fibers, as well as for different ROIs and software versions. The effect of intermediate probe-fiber alignments was also analyzed. Finally, the impact of confounding factors on SWV reproducibility was minimized with a second force-controlled volunteer study, in which the calf was isometrically loaded, and fiber orientation and ROI were well-defined. 2046 in vivoSWE images were acquired to analyze SWV reproducibility with different confounder settings. In healthy volunteers, the main variance-contributing factors were in order of importance muscle tension, fiber orientation, horizontal ROI size and insertion depth. Regression analysis showed significantly reduced SWV with increasing insertion depth for each study material. Parallel probe-fiber orientation, muscle stretch and increasing horizontal ROI size led to significantly higher SWV. Based on the results of the study, we provide recommendations to minimize the impact of confounders in musculoskeletal elastography and discuss the main confounding mechanisms and trade-offs between confounding variables. Coefficients of variation can be significantly reduced with a controlled protocol, if the confounders are clinically taken into account.

Statistics

Citations

Dimensions.ai Metrics
3 citations in Web of Science®
3 citations in Scopus®
Google Scholar™

Altmetrics

Downloads

0 downloads since deposited on 21 Aug 2019
0 downloads since 12 months

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Clinic for Diagnostic and Interventional Radiology
Dewey Decimal Classification:610 Medicine & health
Scopus Subject Areas:Life Sciences > Biophysics
Health Sciences > Radiological and Ultrasound Technology
Physical Sciences > Acoustics and Ultrasonics
Language:English
Date:1 October 2019
Deposited On:21 Aug 2019 10:09
Last Modified:29 Jul 2020 11:11
Publisher:Elsevier
ISSN:0301-5629
OA Status:Closed
Publisher DOI:https://doi.org/10.1016/j.ultrasmedbio.2019.06.417
PubMed ID:31375216

Download

Closed Access: Download allowed only for UZH members