Magnetic resonance imaging of patients with epicardial leads: in vitro evaluation of temperature changes at the lead tip


Balmer, Christian; Gass, Matthias; Dave, Hitendu; Duru, Firat; Luechinger, Roger (2019). Magnetic resonance imaging of patients with epicardial leads: in vitro evaluation of temperature changes at the lead tip. Journal of Interventional Cardiac Electrophysiology, 56(3):321-326.

Abstract

PURPOSE
The aim of this study was to systematically investigate the potential heating effects of magnetic resonance imaging (MRI) in the presence of epicardial leads, which are connected to a device or abandoned, using a series of in vitro measurements.
METHODS
The heating effects of MRI in a 1.5-T scanner were measured at the lead tip in a gel-filled tank. First, a transvenous lead (5086-45 cm, Medtronic) was compared with an epicardial lead (4968-35 cm, Medtronic) with and without connection to an MR-conditional pacemaker. Then, experiments were conducted using various lengths of epicardial leads exposed to MRI.
RESULTS
(1) A temperature rise of + 2.5 °C was observed for the transvenous lead attached to an MRI-conditional pacemaker. The epicardial lead attached to the same pacemaker showed four times higher heating. (2) The transvenous lead without pacemaker showed four times higher heating, and the epicardial lead without pacemaker showed 30 times higher heating. (3) The epicardial lead coiled to 20 cm length without pacemaker showed 9 times higher heating. (4) Experiments with various lengths of epicardial leads showed that the shorter the leads were, the smaller was the heating effect.
CONCLUSION
Standard clinical MRI investigations may result in pronounced heating at the tip of epicardial electrodes. Epicardial leads, which are not connected to a pacemaker and thus mimicking abandoned leads, may even result in a more pronounced rise in temperature at the lead tip. Therefore, current epicardial pacing systems may carry a substantial risk of inducing thermal damage of the neighboring tissue during MRI scanning.

Abstract

PURPOSE
The aim of this study was to systematically investigate the potential heating effects of magnetic resonance imaging (MRI) in the presence of epicardial leads, which are connected to a device or abandoned, using a series of in vitro measurements.
METHODS
The heating effects of MRI in a 1.5-T scanner were measured at the lead tip in a gel-filled tank. First, a transvenous lead (5086-45 cm, Medtronic) was compared with an epicardial lead (4968-35 cm, Medtronic) with and without connection to an MR-conditional pacemaker. Then, experiments were conducted using various lengths of epicardial leads exposed to MRI.
RESULTS
(1) A temperature rise of + 2.5 °C was observed for the transvenous lead attached to an MRI-conditional pacemaker. The epicardial lead attached to the same pacemaker showed four times higher heating. (2) The transvenous lead without pacemaker showed four times higher heating, and the epicardial lead without pacemaker showed 30 times higher heating. (3) The epicardial lead coiled to 20 cm length without pacemaker showed 9 times higher heating. (4) Experiments with various lengths of epicardial leads showed that the shorter the leads were, the smaller was the heating effect.
CONCLUSION
Standard clinical MRI investigations may result in pronounced heating at the tip of epicardial electrodes. Epicardial leads, which are not connected to a pacemaker and thus mimicking abandoned leads, may even result in a more pronounced rise in temperature at the lead tip. Therefore, current epicardial pacing systems may carry a substantial risk of inducing thermal damage of the neighboring tissue during MRI scanning.

Statistics

Citations

Altmetrics

Downloads

1 download since deposited on 12 Dec 2019
1 download since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Clinic for Vascular Surgery
04 Faculty of Medicine > University Children's Hospital Zurich > Medical Clinic
04 Faculty of Medicine > University Hospital Zurich > Clinic for Cardiology
04 Faculty of Medicine > Institute of Biomedical Engineering
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:1 December 2019
Deposited On:12 Dec 2019 10:34
Last Modified:29 Feb 2020 08:25
Publisher:Springer
ISSN:1383-875X
OA Status:Closed
Publisher DOI:https://doi.org/10.1007/s10840-019-00627-7
PubMed ID:31713219

Download

Closed Access: Download allowed only for UZH members