6-Month outcomes of Tricuspid Valve reconstruction for patients with severe tricuspid regurgitation


Abstract

BACKGROUND
Severe tricuspid regurgitation (TR) is associated with high morbidity and mortality rates with limited treatment options.
OBJECTIVES
The authors report the 6-month safety and performance of a transcatheter tricuspid valve reconstruction system in the treatment of moderate to severe functional TR in 30 patients enrolled in the TRI-REPAIR (TrIcuspid Regurgitation RePAIr With CaRdioband Transcatheter System) study.
METHODS
Between October 2016 and July 2017, 30 patients were enrolled in this single-arm, multicenter, prospective trial. Patients were diagnosed with moderate to severe, symptomatic TR in the absence of untreated left-heart disease and deemed inoperable because of unacceptable risk for open-heart surgery by the local heart team. Clinical, functional, and echocardiographic data were prospectively collected before and up to 6 months post-procedure. An independent core lab assessed all echocardiographic data, and an independent clinical event committee adjudicated the safety events.
RESULTS
Mean patient age was 75 years, 73% were female, and 23% had ischemic heart disease. At baseline, 83% were in New York Heart Association (NYHA) functional class III to IV, and mean left ventricular ejection fraction was 58%. Technical success was 100%. Through 6 months, 3 patients died. Between 6 months and baseline, echocardiography showed average reductions of annular septolateral diameter of 9% (42 mm vs. 38 mm; p < 0.01), proximal isovelocity surface area effective regurgitant orifice area of 50% (0.8 cm$^{2}$ vs. 0.4 cm$^{2}$; p < 0.01), and mean vena contracta width of 28% (1.2 cm vs. 0.9 cm; p < 0.01). Clinical assessment showed that 76% of patients improved by at least 1 NYHA functional class with 88% in NYHA functional class I or II. Six-minute walk distance improved by 60 m (p < 0.01), and Kansas City Cardiomyopathy Questionnaire score improved by 24 points (p < 0.01).
CONCLUSIONS
Six-month outcomes show that the system performs as intended and appears to be safe in patients with symptomatic and moderate to severe functional TR. Significant reduction of TR through decrease of annular dimensions, improvements in heart failure symptoms, quality of life, and exercise capacity were observed. Further studies are warranted to validate these initial promising results. (TrIcuspid Regurgitation RePAIr With CaRdioband Transcatheter System [TRI-REPAIR]; NCT02981953).

Abstract

BACKGROUND
Severe tricuspid regurgitation (TR) is associated with high morbidity and mortality rates with limited treatment options.
OBJECTIVES
The authors report the 6-month safety and performance of a transcatheter tricuspid valve reconstruction system in the treatment of moderate to severe functional TR in 30 patients enrolled in the TRI-REPAIR (TrIcuspid Regurgitation RePAIr With CaRdioband Transcatheter System) study.
METHODS
Between October 2016 and July 2017, 30 patients were enrolled in this single-arm, multicenter, prospective trial. Patients were diagnosed with moderate to severe, symptomatic TR in the absence of untreated left-heart disease and deemed inoperable because of unacceptable risk for open-heart surgery by the local heart team. Clinical, functional, and echocardiographic data were prospectively collected before and up to 6 months post-procedure. An independent core lab assessed all echocardiographic data, and an independent clinical event committee adjudicated the safety events.
RESULTS
Mean patient age was 75 years, 73% were female, and 23% had ischemic heart disease. At baseline, 83% were in New York Heart Association (NYHA) functional class III to IV, and mean left ventricular ejection fraction was 58%. Technical success was 100%. Through 6 months, 3 patients died. Between 6 months and baseline, echocardiography showed average reductions of annular septolateral diameter of 9% (42 mm vs. 38 mm; p < 0.01), proximal isovelocity surface area effective regurgitant orifice area of 50% (0.8 cm$^{2}$ vs. 0.4 cm$^{2}$; p < 0.01), and mean vena contracta width of 28% (1.2 cm vs. 0.9 cm; p < 0.01). Clinical assessment showed that 76% of patients improved by at least 1 NYHA functional class with 88% in NYHA functional class I or II. Six-minute walk distance improved by 60 m (p < 0.01), and Kansas City Cardiomyopathy Questionnaire score improved by 24 points (p < 0.01).
CONCLUSIONS
Six-month outcomes show that the system performs as intended and appears to be safe in patients with symptomatic and moderate to severe functional TR. Significant reduction of TR through decrease of annular dimensions, improvements in heart failure symptoms, quality of life, and exercise capacity were observed. Further studies are warranted to validate these initial promising results. (TrIcuspid Regurgitation RePAIr With CaRdioband Transcatheter System [TRI-REPAIR]; NCT02981953).

Statistics

Citations

Dimensions.ai Metrics
15 citations in Web of Science®
11 citations in Scopus®
Google Scholar™

Altmetrics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Clinic for Vascular Surgery
Dewey Decimal Classification:610 Medicine & health
Uncontrolled Keywords:TRI-REPAIR annular reduction tricuspid regurgitation tricuspid repair
Language:English
Date:23 April 2019
Deposited On:10 Jan 2020 06:57
Last Modified:10 Jan 2020 07:13
Publisher:Elsevier
ISSN:0735-1097
OA Status:Closed
Free access at:Publisher DOI. An embargo period may apply.
Publisher DOI:https://doi.org/10.1016/j.jacc.2019.01.062
PubMed ID:30999993

Download

Full text not available from this repository.
View at publisher

Get full-text in a library