Blood trauma potential of the HeartWare Ventricular Assist Device in pediatric patients


Granegger, Marcus; Thamsen, Bente; Schlöglhofer, Thomas; Lach, Selina; Escher, Andreas; Haas, Thorsten; Meboldt, Mirko; Schweiger, Martin; Hübler, Michael; Zimpfer, Daniel (2020). Blood trauma potential of the HeartWare Ventricular Assist Device in pediatric patients. Journal of Thoracic and Cardiovascular Surgery, 159(4):1519-1527.e1.

Abstract

OBJECTIVE

Mechanical circulatory support has become a standard therapy for adult patients with end-stage heart failure. For pediatric patients, technologic development lags behind with no currently approved implantable rotary blood pump. As an alternative, the HeartWare Ventricular Assist Device (Medtronic, Minneapolis, Minn), originally designed for adults, is increasingly used in pediatric patients. The aim of this multicenter study was to assess in silico, in vitro, and in vivo the blood trauma potential of this pump in pediatric application.

METHODS

Clinical outcome and indicators for in vivo blood trauma were investigated retrospectively in 14 pediatric patients with the HeartWare Ventricular Assist Device (age 11.3 ± 4.8 years). Blood trauma mechanisms of the HeartWare Ventricular Assist Device were examined in silico and in vitro at an adult and pediatric operating point (5 L/min and 2.5 L/min at 2800 rpm and 2200 rpm, respectively). The flow was simulated by computational fluid dynamics and analyzed regarding flow structures, shear stresses, and washout. Hemolysis was assessed with pumps circulating bovine blood in a temperate flow circuit.

RESULTS

In the retrospective in vivo analysis, lactate dehydrogenase and D-dimer values were 1.5- and 3-fold elevated, respectively, compared with adult patients with the HeartWare Ventricular Assist Device. Major bleedings were observed in 42.9%, and suspected pump thrombosis and neurologic dysfunction were observed in 14.3% of all patients. In the pediatric conditions, simulations predicted elevated mechanical stress profile below 50 Pa, more stagnant flow field, and longer washout times within the pump. In vitro measurements revealed an increased normalized index of hemolysis (17.5 vs 8.2 mg/100 L; P = .0021).

CONCLUSIONS

The HeartWare Ventricular Assist Device, operated at lower speeds and flows, induces elevated blood trauma. Further studies are required to assess the clinical implications of these findings.

Abstract

OBJECTIVE

Mechanical circulatory support has become a standard therapy for adult patients with end-stage heart failure. For pediatric patients, technologic development lags behind with no currently approved implantable rotary blood pump. As an alternative, the HeartWare Ventricular Assist Device (Medtronic, Minneapolis, Minn), originally designed for adults, is increasingly used in pediatric patients. The aim of this multicenter study was to assess in silico, in vitro, and in vivo the blood trauma potential of this pump in pediatric application.

METHODS

Clinical outcome and indicators for in vivo blood trauma were investigated retrospectively in 14 pediatric patients with the HeartWare Ventricular Assist Device (age 11.3 ± 4.8 years). Blood trauma mechanisms of the HeartWare Ventricular Assist Device were examined in silico and in vitro at an adult and pediatric operating point (5 L/min and 2.5 L/min at 2800 rpm and 2200 rpm, respectively). The flow was simulated by computational fluid dynamics and analyzed regarding flow structures, shear stresses, and washout. Hemolysis was assessed with pumps circulating bovine blood in a temperate flow circuit.

RESULTS

In the retrospective in vivo analysis, lactate dehydrogenase and D-dimer values were 1.5- and 3-fold elevated, respectively, compared with adult patients with the HeartWare Ventricular Assist Device. Major bleedings were observed in 42.9%, and suspected pump thrombosis and neurologic dysfunction were observed in 14.3% of all patients. In the pediatric conditions, simulations predicted elevated mechanical stress profile below 50 Pa, more stagnant flow field, and longer washout times within the pump. In vitro measurements revealed an increased normalized index of hemolysis (17.5 vs 8.2 mg/100 L; P = .0021).

CONCLUSIONS

The HeartWare Ventricular Assist Device, operated at lower speeds and flows, induces elevated blood trauma. Further studies are required to assess the clinical implications of these findings.

Statistics

Citations

Dimensions.ai Metrics
19 citations in Web of Science®
20 citations in Scopus®
Google Scholar™

Altmetrics

Downloads

0 downloads since deposited on 09 Jan 2020
0 downloads since 12 months

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Children's Hospital Zurich > Clinic for Surgery
Dewey Decimal Classification:610 Medicine & health
Scopus Subject Areas:Health Sciences > Surgery
Health Sciences > Pulmonary and Respiratory Medicine
Health Sciences > Cardiology and Cardiovascular Medicine
Language:English
Date:1 April 2020
Deposited On:09 Jan 2020 11:15
Last Modified:26 Jan 2022 23:28
Publisher:Elsevier
ISSN:0022-5223
OA Status:Closed
Free access at:Publisher DOI. An embargo period may apply.
Publisher DOI:https://doi.org/10.1016/j.jtcvs.2019.06.084
PubMed ID:31444074