Fracture resistance and 2-body wear of 3-dimensional–printed occlusal devices


Lutz, Anna-Maria; Hampe, Rüdiger; Roos, Malgorzata; Lümkemann, Nina; Eichberger, Marlis; Stawarczyk, Bogna (2019). Fracture resistance and 2-body wear of 3-dimensional–printed occlusal devices. Journal of Prosthetic Dentistry, 121(1):166-172.

Abstract

Statement of problem

Polymeric material for 3-dimensional printing can be used to fabricate occlusal devices. However, information about fracture resistance and wear is scarce.
Purpose

The purpose of this in vitro study was to investigate the fracture resistance and 2-body wear of 3-dimensional–printed (3DP) (FotoDent splint; Dreve Dentamid GmbH), milled polymethylmethacrylate (CAM) (Temp Basic; Transpa 95H16, Zirkonzahn GmbH), and conventionally fabricated polymethylmethacrylate (CAST) (Castdon; Dreve Dentamid GmbH) occlusal devices.
Material and Methods

A total of 96 occlusal devices were prepared according to the 3 different manufacturing techniques 3DP, CAM, and CAST (n=32). For each manufacturing technique, specimens were further divided into initial fracture resistance tests (n=16) and artificial aging in the mastication simulator (50 N, 37°C) with 2-body wear followed by fracture resistance tests (n=16). The fracture resistance was determined using a universal testing machine (1 mm/min). The wear was measured after 20 000 and 120 000 mastication cycles with the replica technique, mapped with a laser scanner, and quantified in R software. Data were analyzed using a 2-way ANOVA followed by a 1-way ANOVA with Scheffé or Games-Howell post hoc tests, repeated measures ANOVA with corrected Greenhouse-Geisser P values, and the Levene, Mann-Whitney, and paired t tests (α=.05).
Results

CAM presented higher initial fracture resistance than 3DP or CAST (P<.001). After mastication simulation, CAM followed by 3DP showed higher fracture resistance than CAST (P<.001). Mastication simulation decreased the fracture resistance for CAM and CAST (P<.001) but not for 3DP (P=.78). Three-dimensional–printed occlusal devices showed the highest material volume loss, followed by CAM and the lowest in CAST (P<.001).
Conclusions

Three-dimensional–printed occlusal devices showed lower wear resistance and lower fracture resistance than those milled or conventionally fabricated. Therefore, only short-term application in the mouth is recommended. Further developments of occlusal device material for 3-dimensional printing are necessary.

Abstract

Statement of problem

Polymeric material for 3-dimensional printing can be used to fabricate occlusal devices. However, information about fracture resistance and wear is scarce.
Purpose

The purpose of this in vitro study was to investigate the fracture resistance and 2-body wear of 3-dimensional–printed (3DP) (FotoDent splint; Dreve Dentamid GmbH), milled polymethylmethacrylate (CAM) (Temp Basic; Transpa 95H16, Zirkonzahn GmbH), and conventionally fabricated polymethylmethacrylate (CAST) (Castdon; Dreve Dentamid GmbH) occlusal devices.
Material and Methods

A total of 96 occlusal devices were prepared according to the 3 different manufacturing techniques 3DP, CAM, and CAST (n=32). For each manufacturing technique, specimens were further divided into initial fracture resistance tests (n=16) and artificial aging in the mastication simulator (50 N, 37°C) with 2-body wear followed by fracture resistance tests (n=16). The fracture resistance was determined using a universal testing machine (1 mm/min). The wear was measured after 20 000 and 120 000 mastication cycles with the replica technique, mapped with a laser scanner, and quantified in R software. Data were analyzed using a 2-way ANOVA followed by a 1-way ANOVA with Scheffé or Games-Howell post hoc tests, repeated measures ANOVA with corrected Greenhouse-Geisser P values, and the Levene, Mann-Whitney, and paired t tests (α=.05).
Results

CAM presented higher initial fracture resistance than 3DP or CAST (P<.001). After mastication simulation, CAM followed by 3DP showed higher fracture resistance than CAST (P<.001). Mastication simulation decreased the fracture resistance for CAM and CAST (P<.001) but not for 3DP (P=.78). Three-dimensional–printed occlusal devices showed the highest material volume loss, followed by CAM and the lowest in CAST (P<.001).
Conclusions

Three-dimensional–printed occlusal devices showed lower wear resistance and lower fracture resistance than those milled or conventionally fabricated. Therefore, only short-term application in the mouth is recommended. Further developments of occlusal device material for 3-dimensional printing are necessary.

Statistics

Citations

Dimensions.ai Metrics

Altmetrics

Downloads

1 download since deposited on 15 Jan 2020
1 download since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > Epidemiology, Biostatistics and Prevention Institute (EBPI)
Dewey Decimal Classification:610 Medicine & health
Uncontrolled Keywords:Oral Surgery
Language:English
Date:1 January 2019
Deposited On:15 Jan 2020 12:31
Last Modified:19 Jan 2020 07:05
Publisher:Elsevier
ISSN:0022-3913
OA Status:Closed
Publisher DOI:https://doi.org/10.1016/j.prosdent.2018.04.007
PubMed ID:30647000

Download

Closed Access: Download allowed only for UZH members