Radiation Dose to the Fetus From Computed Tomography of Pregnant Patients-Development and Validation of a Web-Based Tool


Saltybaeva, Natalia; Platon, Alexandra; Poletti, Pierre-Alexandre; Hinzpeter, Ricarda; Merce, Marta Sans; Alkadhi, Hatem (2020). Radiation Dose to the Fetus From Computed Tomography of Pregnant Patients-Development and Validation of a Web-Based Tool. Investigative Radiology, Publish :Epub ahead of print.

Abstract

OBJECTIVE

Estimations of radiation dose absorbed by the fetus from computed tomography (CT) in pregnant patients is mandatory, but currently available methods are not feasible in clinical routine. The aims of this study were to develop and validate a tool for assessment of fetal dose from CT of pregnant patients and to develop a user-friendly web interface for fast fetal dose calculations.

METHODS

In the first study part, 750 Monte Carlo (MC) simulations were performed on phantoms representing pregnant patients at various gestational stages. The MC code simulating vendor-independent dose distributions was validated against CT dose index (CTDI) measurements performed on CT scanners of 2 vendors. The volume CTDI-normalized fetal dose values from MC simulations were used for developing the computational algorithm enabling fetal dose assessments from CT of various body regions at different exposure settings. In the institutional review board-approved second part, the algorithm was validated against patient-specific MC simulations performed on CT data of 29 pregnant patients (gestational ages 8-35 weeks) who underwent CT. Furthermore, the tool was compared with a commercially available software. A user-friendly web-based interface for fetal dose calculations was created.

RESULTS

Weighted CTDI values obtained from MC simulations were in excellent agreement with measurements performed on the 2 CT systems (average error, 4%). The median fetal dose from abdominal CT in pregnant patients was 2.7 mGy, showing moderate correlation with maternal perimeter (r = 0.69). The algorithm provided accurate estimates of fetal doses (average error, 11%), being more accurate than the commercially available tool. The web-based interface (www.fetaldose.org) enabling vendor-independent calculations of fetal doses from CT requires the input of gestational age, volume CTDI, tube voltage, and scan region.

CONCLUSIONS

A tool for fetal dose assessments from CT of pregnant patients was developed and validated being freely available on a user-friendly web interface.

Abstract

OBJECTIVE

Estimations of radiation dose absorbed by the fetus from computed tomography (CT) in pregnant patients is mandatory, but currently available methods are not feasible in clinical routine. The aims of this study were to develop and validate a tool for assessment of fetal dose from CT of pregnant patients and to develop a user-friendly web interface for fast fetal dose calculations.

METHODS

In the first study part, 750 Monte Carlo (MC) simulations were performed on phantoms representing pregnant patients at various gestational stages. The MC code simulating vendor-independent dose distributions was validated against CT dose index (CTDI) measurements performed on CT scanners of 2 vendors. The volume CTDI-normalized fetal dose values from MC simulations were used for developing the computational algorithm enabling fetal dose assessments from CT of various body regions at different exposure settings. In the institutional review board-approved second part, the algorithm was validated against patient-specific MC simulations performed on CT data of 29 pregnant patients (gestational ages 8-35 weeks) who underwent CT. Furthermore, the tool was compared with a commercially available software. A user-friendly web-based interface for fetal dose calculations was created.

RESULTS

Weighted CTDI values obtained from MC simulations were in excellent agreement with measurements performed on the 2 CT systems (average error, 4%). The median fetal dose from abdominal CT in pregnant patients was 2.7 mGy, showing moderate correlation with maternal perimeter (r = 0.69). The algorithm provided accurate estimates of fetal doses (average error, 11%), being more accurate than the commercially available tool. The web-based interface (www.fetaldose.org) enabling vendor-independent calculations of fetal doses from CT requires the input of gestational age, volume CTDI, tube voltage, and scan region.

CONCLUSIONS

A tool for fetal dose assessments from CT of pregnant patients was developed and validated being freely available on a user-friendly web interface.

Statistics

Citations

Dimensions.ai Metrics

Altmetrics

Downloads

0 downloads since deposited on 13 Jul 2020
0 downloads since 12 months

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Clinic for Diagnostic and Interventional Radiology
Dewey Decimal Classification:610 Medicine & health
Language:English
Date:22 June 2020
Deposited On:13 Jul 2020 10:40
Last Modified:29 Jul 2020 15:25
Publisher:Lippincott Williams & Wilkins
ISSN:0020-9996
OA Status:Closed
Publisher DOI:https://doi.org/10.1097/RLI.0000000000000701
PubMed ID:32604386

Download

Closed Access: Download allowed only for UZH members