Validation of deep-learning image reconstruction for coronary computed tomography angiography: Impact on noise, image quality and diagnostic accuracy


Benz, Dominik C; Benetos, Georgios; Rampidis, Georgios; von Felten, Elia; Bakula, Adam; Sustar, Aleksandra; Kudura, Ken; Messerli, Michael; Fuchs, Tobias A; Gebhard, Catherine; Pazhenkottil, Aju P; Kaufmann, Philipp A; Buechel, Ronny R (2020). Validation of deep-learning image reconstruction for coronary computed tomography angiography: Impact on noise, image quality and diagnostic accuracy. Journal of Cardiovascular Computed Tomography, 14(5):444-451.

Abstract

BACKGROUND

Advances in image reconstruction are necessary to decrease radiation exposure from coronary CT angiography (CCTA) further, but iterative reconstruction has been shown to degrade image quality at high levels. Deep-learning image reconstruction (DLIR) offers unique opportunities to overcome these limitations. The present study compared the impact of DLIR and adaptive statistical iterative reconstruction-Veo (ASiR-V) on quantitative and qualitative image parameters and the diagnostic accuracy of CCTA using invasive coronary angiography (ICA) as the standard of reference.

METHODS

This retrospective study includes 43 patients who underwent clinically indicated CCTA and ICA. Datasets were reconstructed with ASiR-V 70% (using standard [SD] and high-definition [HD] kernels) and with DLIR at different levels (i.e., medium [M] and high [H]). Image noise, image quality, and coronary luminal narrowing were evaluated by three blinded readers. Diagnostic accuracy was compared against ICA.

RESULTS

Noise did not significantly differ between ASiR-V SD and DLIR-M (37 vs. 37 HU, p = 1.000), but was significantly lower in DLIR-H (30 HU, p < 0.001) and higher in ASiR-V HD (53 HU, p < 0.001). Image quality was higher for DLIR-M and DLIR-H (3.4-3.8 and 4.2-4.6) compared to ASiR-V SD and HD (2.1-2.7 and 1.8-2.2; p < 0.001), with DLIR-H yielding the highest image quality. Consistently across readers, no significant differences in sensitivity (88% vs. 92%; p = 0.453), specificity (73% vs. 73%; p = 0.583) and diagnostic accuracy (80% vs. 82%; p = 0.366) were found between ASiR-V HD and DLIR-H.

CONCLUSION

DLIR significantly reduces noise in CCTA compared to ASiR-V, while yielding superior image quality at equal diagnostic accuracy.

Abstract

BACKGROUND

Advances in image reconstruction are necessary to decrease radiation exposure from coronary CT angiography (CCTA) further, but iterative reconstruction has been shown to degrade image quality at high levels. Deep-learning image reconstruction (DLIR) offers unique opportunities to overcome these limitations. The present study compared the impact of DLIR and adaptive statistical iterative reconstruction-Veo (ASiR-V) on quantitative and qualitative image parameters and the diagnostic accuracy of CCTA using invasive coronary angiography (ICA) as the standard of reference.

METHODS

This retrospective study includes 43 patients who underwent clinically indicated CCTA and ICA. Datasets were reconstructed with ASiR-V 70% (using standard [SD] and high-definition [HD] kernels) and with DLIR at different levels (i.e., medium [M] and high [H]). Image noise, image quality, and coronary luminal narrowing were evaluated by three blinded readers. Diagnostic accuracy was compared against ICA.

RESULTS

Noise did not significantly differ between ASiR-V SD and DLIR-M (37 vs. 37 HU, p = 1.000), but was significantly lower in DLIR-H (30 HU, p < 0.001) and higher in ASiR-V HD (53 HU, p < 0.001). Image quality was higher for DLIR-M and DLIR-H (3.4-3.8 and 4.2-4.6) compared to ASiR-V SD and HD (2.1-2.7 and 1.8-2.2; p < 0.001), with DLIR-H yielding the highest image quality. Consistently across readers, no significant differences in sensitivity (88% vs. 92%; p = 0.453), specificity (73% vs. 73%; p = 0.583) and diagnostic accuracy (80% vs. 82%; p = 0.366) were found between ASiR-V HD and DLIR-H.

CONCLUSION

DLIR significantly reduces noise in CCTA compared to ASiR-V, while yielding superior image quality at equal diagnostic accuracy.

Statistics

Citations

Dimensions.ai Metrics
88 citations in Web of Science®
98 citations in Scopus®
Google Scholar™

Altmetrics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Clinic for Nuclear Medicine
Dewey Decimal Classification:610 Medicine & health
Scopus Subject Areas:Health Sciences > Radiology, Nuclear Medicine and Imaging
Health Sciences > Cardiology and Cardiovascular Medicine
Language:English
Date:1 September 2020
Deposited On:03 Sep 2020 09:30
Last Modified:24 Nov 2023 02:40
Publisher:Elsevier
ISSN:1876-861X
OA Status:Closed
Publisher DOI:https://doi.org/10.1016/j.jcct.2020.01.002
PubMed ID:31974008
Full text not available from this repository.