Early diagnosis of myocardial infarction with point-of-care high-sensitivity cardiac troponin I.


Abstract

BACKGROUND: Until now, high-sensitivity cardiac troponin (hs-cTn) assays were mainly developed for large central laboratory platforms.
OBJECTIVES: This study aimed to assess the clinical performance of a point-of-care (POC)-hs-cTnI assay in patients with suspected myocardial infarction (MI).
METHODS: This study enrolled patients presenting to the emergency department with symptoms suggestive of MI. Two cardiologists centrally adjudicated the final diagnosis using all clinical data including cardiac imaging. The primary objective was to directly compare diagnostic accuracy of POC-hs-cTnI-TriageTrue versus best-validated central laboratory assays. Secondary objectives included the derivation and validation of a POC-hs-cTnI-TriageTrue-specific 0/1-h algorithm.
RESULTS: MI was the adjudicated final diagnosis in 178 of 1,261 patients (14%). The area under the curve (AUC) for POC-hs-cTnI-TriageTrue at presentation was 0.95 (95% confidence interval [CI]: 0.93 to 0.96) and was at least comparable to hs-cTnT-Elecsys (AUC: 0.94; 95% CI: 0.93 to 0.96; p = 0.213) and hs-cTnI-Architect (AUC: 0.92; 95% CI: 0.90 to 0.93; p < 0.001). A single cutoff concentration <3 ng/l at presentation identified 45% of patients at low risk with a negative predictive value (NPV) of 100% (95% CI: 99.4% to 100%). A single cutoff concentration >60 ng/l identified patients at high risk with a positive predictive value (PPV) of 76.8% (95% CI: 68.9% to 83.6%). The 0/1-h algorithm ruled out 55% of patients (NPV: 100%; 95% CI: 98.8% to 100%), and ruled in 18% of patients (PPV: 76.8%; 95% CI: 67.2% to 84.7%). Ruled-out patients had cumulative event rates of 0% at 30 days and 1.6% at 2 years. This study confirmed these findings in a secondary analysis including hs-cTnI-Architect for central adjudication.
CONCLUSIONS: The POC-hs-cTnI-TriageTrue assay provides high diagnostic accuracy in patients with suspected MI with a clinical performance that is at least comparable to that of best-validated central laboratory assays. (Advantageous Predictors of Acute Coronary Syndromes Evaluation Study [APACE]; NCT00470587).

Abstract

BACKGROUND: Until now, high-sensitivity cardiac troponin (hs-cTn) assays were mainly developed for large central laboratory platforms.
OBJECTIVES: This study aimed to assess the clinical performance of a point-of-care (POC)-hs-cTnI assay in patients with suspected myocardial infarction (MI).
METHODS: This study enrolled patients presenting to the emergency department with symptoms suggestive of MI. Two cardiologists centrally adjudicated the final diagnosis using all clinical data including cardiac imaging. The primary objective was to directly compare diagnostic accuracy of POC-hs-cTnI-TriageTrue versus best-validated central laboratory assays. Secondary objectives included the derivation and validation of a POC-hs-cTnI-TriageTrue-specific 0/1-h algorithm.
RESULTS: MI was the adjudicated final diagnosis in 178 of 1,261 patients (14%). The area under the curve (AUC) for POC-hs-cTnI-TriageTrue at presentation was 0.95 (95% confidence interval [CI]: 0.93 to 0.96) and was at least comparable to hs-cTnT-Elecsys (AUC: 0.94; 95% CI: 0.93 to 0.96; p = 0.213) and hs-cTnI-Architect (AUC: 0.92; 95% CI: 0.90 to 0.93; p < 0.001). A single cutoff concentration <3 ng/l at presentation identified 45% of patients at low risk with a negative predictive value (NPV) of 100% (95% CI: 99.4% to 100%). A single cutoff concentration >60 ng/l identified patients at high risk with a positive predictive value (PPV) of 76.8% (95% CI: 68.9% to 83.6%). The 0/1-h algorithm ruled out 55% of patients (NPV: 100%; 95% CI: 98.8% to 100%), and ruled in 18% of patients (PPV: 76.8%; 95% CI: 67.2% to 84.7%). Ruled-out patients had cumulative event rates of 0% at 30 days and 1.6% at 2 years. This study confirmed these findings in a secondary analysis including hs-cTnI-Architect for central adjudication.
CONCLUSIONS: The POC-hs-cTnI-TriageTrue assay provides high diagnostic accuracy in patients with suspected MI with a clinical performance that is at least comparable to that of best-validated central laboratory assays. (Advantageous Predictors of Acute Coronary Syndromes Evaluation Study [APACE]; NCT00470587).

Statistics

Citations

Dimensions.ai Metrics
2 citations in Web of Science®
3 citations in Scopus®
Google Scholar™

Altmetrics

Downloads

0 downloads since deposited on 16 Oct 2020
0 downloads since 12 months

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Institute of Clinical Chemistry
Dewey Decimal Classification:610 Medicine & health
540 Chemistry
Scopus Subject Areas:Health Sciences > Cardiology and Cardiovascular Medicine
Language:English
Date:17 March 2020
Deposited On:16 Oct 2020 15:07
Last Modified:18 Oct 2020 08:34
Publisher:Elsevier
ISSN:0735-1097
OA Status:Hybrid
Free access at:Publisher DOI. An embargo period may apply.
Publisher DOI:https://doi.org/10.1016/j.jacc.2019.12.065
PubMed ID:32164884

Download

Hybrid Open Access

Download PDF  'Early diagnosis of myocardial infarction with point-of-care high-sensitivity cardiac troponin I.'.
Preview
Content: Published Version
Language: English
Filetype: PDF
Size: 1MB
View at publisher
Licence: Creative Commons: Attribution-NonCommercial-NoDerivatives 4.0 International (CC BY-NC-ND 4.0)