Performance characteristics of the new Eurosets magnetically suspended centrifugal pump


Berdajs, Denis; von Segesser, Ludwig K; Maisano, Francesco; Milano, Guiseppina; Ferrari, Enrico (2020). Performance characteristics of the new Eurosets magnetically suspended centrifugal pump. Perfusion:Epub ahead of print.

Abstract

OBJECTIVE
The aim was to evaluate the performance of a newly developed magnetically suspended centrifugal pump head intended for use as a ventricular assistance device with a newly developed extracorporeal membrane oxygenator setup.
METHODS
In an experimental setup, an extracorporeal membrane oxygenator circuit was established in three calves with a mean weight of 68.2 ± 2.0 kg. A magnetically levitated centrifugal pump was tested, along with a newly designed extracorporeal membrane oxygenator console, at three different flow ranges: (a) 0.0 to 5.2 L/min, (b) 0.0 to 7.1 L/min, and (c) 0.0 to 6.0 L/min. For each setup, the animals were supported by a circuit for 6 h. Blood samples were collected just before caridiopulmonary bypass (CPB) after 10 min on bypass and after 1, 2, 5, and 6 h of perfusion for hemolysis determination and biochemical tests. Values were recorded for blood pressure, mean flow, and pump rotational speed. Analysis of variance was used for repeated measurements.
RESULTS
Mean pump flows achieved during the three 6 h pump runs for the three pump heads studied were as follows: (a) flow range 0.0 to 5.2 L/min, 3.6 ± 1.5 L/min, (b) flow range 0.0 to 7.1 L/min, 4.9 ± 1.3 L/min, and (c) flow range 0.0 to 6.0 L/min, 3.8 ± 1.5 L/min. Blood trauma, evaluated by plasma hemoglobin and lactate dehydrogenase levels, did not help in detecting any significant hemolysis. Thrombocytes and white blood cell count profiles showed no significant differences between the groups at the end of the 6 h perfusion. At the end of testing, no clot deposition was found in the oxygenator, and there was no evidence of peripheral emboli.
CONCLUSION
The results suggest that the newly developed magnetically suspended centrifugal pump head provides satisfactory hydrodynamic performance in an acute perfusion scenario without increasing hemolysis.

Abstract

OBJECTIVE
The aim was to evaluate the performance of a newly developed magnetically suspended centrifugal pump head intended for use as a ventricular assistance device with a newly developed extracorporeal membrane oxygenator setup.
METHODS
In an experimental setup, an extracorporeal membrane oxygenator circuit was established in three calves with a mean weight of 68.2 ± 2.0 kg. A magnetically levitated centrifugal pump was tested, along with a newly designed extracorporeal membrane oxygenator console, at three different flow ranges: (a) 0.0 to 5.2 L/min, (b) 0.0 to 7.1 L/min, and (c) 0.0 to 6.0 L/min. For each setup, the animals were supported by a circuit for 6 h. Blood samples were collected just before caridiopulmonary bypass (CPB) after 10 min on bypass and after 1, 2, 5, and 6 h of perfusion for hemolysis determination and biochemical tests. Values were recorded for blood pressure, mean flow, and pump rotational speed. Analysis of variance was used for repeated measurements.
RESULTS
Mean pump flows achieved during the three 6 h pump runs for the three pump heads studied were as follows: (a) flow range 0.0 to 5.2 L/min, 3.6 ± 1.5 L/min, (b) flow range 0.0 to 7.1 L/min, 4.9 ± 1.3 L/min, and (c) flow range 0.0 to 6.0 L/min, 3.8 ± 1.5 L/min. Blood trauma, evaluated by plasma hemoglobin and lactate dehydrogenase levels, did not help in detecting any significant hemolysis. Thrombocytes and white blood cell count profiles showed no significant differences between the groups at the end of the 6 h perfusion. At the end of testing, no clot deposition was found in the oxygenator, and there was no evidence of peripheral emboli.
CONCLUSION
The results suggest that the newly developed magnetically suspended centrifugal pump head provides satisfactory hydrodynamic performance in an acute perfusion scenario without increasing hemolysis.

Statistics

Citations

Altmetrics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Clinic for Cardiac Surgery
04 Faculty of Medicine > Cardiocentro Ticino
Dewey Decimal Classification:610 Medicine & health
Scopus Subject Areas:Health Sciences > Radiology, Nuclear Medicine and Imaging
Social Sciences & Humanities > Safety Research
Health Sciences > Cardiology and Cardiovascular Medicine
Health Sciences > Advanced and Specialized Nursing
Language:English
Date:20 June 2020
Deposited On:13 Jan 2021 06:34
Last Modified:16 Feb 2021 17:22
Publisher:Sage Publications
ISSN:0267-6591
OA Status:Closed
Publisher DOI:https://doi.org/10.1177/0267659120931990
PubMed ID:32567504

Download

Full text not available from this repository.
View at publisher

Get full-text in a library