Navigation auf zora.uzh.ch

Search ZORA

ZORA (Zurich Open Repository and Archive)

Epicardial left atrial appendage occlusion with a new medical device: assessment of procedural feasibility, safety and efficacy in a large animal model

Emmert, Maximilian Y; Firstenberg, Michael S; Martella, Arthur T; Lau, Liming; Zlock, Stephen; Mohan, Ashik; Spangler, Taylor; Currie, Sarah; Salzberg, Sacha P; Caliskan, Etem (2020). Epicardial left atrial appendage occlusion with a new medical device: assessment of procedural feasibility, safety and efficacy in a large animal model. Journal of Cardiothoracic Surgery, 15:56.

Abstract

BACKGROUND

Left atrial appendage occlusion (LAAO) represents a treatment alternative to anticoagulation in patients with atrial fibrillation. We evaluate a novel device for epicardial LAAO in a translational canine model.

METHODS

Nine hounds (n = 9) were used to assess usability, safety, and efficacy of the TigerPaw Pro (TPP) device for epicardial LAAO. Following baseline imaging (intra-cardiac echocardiography (ICE) and angiography) and intraoperative visual inspection, usability was tested via a ``closure/re-opening`` maneuver followed by deployment of a total of twenty TPP devices (n = 20) on the left and right atrial appendages respectively. Procedural safety was evaluated by assessing for adverse-events via direct Epicardial inspection and endocardial imaging. Efficacy evaluation included assessment of device positioning, presence of residual stumps and completeness of closure. Post-mortem evaluation was performed to confirm safety and efficacy.

RESULTS

Usability testing of all TPP devices was successful (n = 20;100%, delivery-time range 22-120 s) without any procedural adverse-events (tissue damage or tears, bleeding, vessel-impingement, structural impact). All devices fully traversed the ostium (n = 18) or appendage body (n = 2), and conformed smoothly to adjacent cardiac anatomy. In nineteen deployments (n = 19;95%), all device connector pairs were fully engaged, while in one TPP device the most distal pair remained unengaged. ICE and post-mortem inspections revealed complete closure of all appendage ostia (n = 18;100%) and only in one case a small residual stump was detected. Intraoperative safety findings were further confirmed post-mortem. Devices created a nearly smooth line of closure via symmetric endocardial tissue-coaptation.

CONCLUSIONS

In this preclinical model, the TPP demonstrated good ease of use for ostial access, ability to re-position (after engagement) and rapid deployment, while achieving safe and effective LAAO.

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > Institute for Regenerative Medicine (IREM)
Dewey Decimal Classification:610 Medicine & health
Scopus Subject Areas:Health Sciences > Surgery
Health Sciences > Pulmonary and Respiratory Medicine
Health Sciences > Cardiology and Cardiovascular Medicine
Language:English
Date:3 April 2020
Deposited On:16 Feb 2021 14:32
Last Modified:11 Mar 2025 04:40
Publisher:BioMed Central
ISSN:1749-8090
OA Status:Gold
Free access at:PubMed ID. An embargo period may apply.
Publisher DOI:https://doi.org/10.1186/s13019-020-01096-0
PubMed ID:32245388
Download PDF  'Epicardial left atrial appendage occlusion with a new medical device: assessment of procedural feasibility, safety and efficacy in a large animal model'.
Preview
  • Content: Published Version
  • Licence: Creative Commons: Attribution 4.0 International (CC BY 4.0)

Metadata Export

Statistics

Citations

Dimensions.ai Metrics
5 citations in Web of Science®
4 citations in Scopus®
Google Scholar™

Altmetrics

Downloads

47 downloads since deposited on 16 Feb 2021
15 downloads since 12 months
Detailed statistics

Authors, Affiliations, Collaborations

Similar Publications