Reduced fracture load of dental implants after implantoplasty with different instrumentation sequences. An in vitro study

Jorio, Isabelle Charlotte. Reduced fracture load of dental implants after implantoplasty with different instrumentation sequences. An in vitro study. 2021, University of Zurich, Faculty of Medicine.

Abstract

Objectives

To assess the mechanical stability of implants after implantoplasty and thermocyclic loading, the residual thickness of the instrumented areas and neighbouring tooth injury due to implantoplasty.
Materials and Methods

Using a phantom head simulator and maxillary model implants were subjected to an implantoplasty procedure. Thirty implants were randomly assigned to receive one of three instrumentation sequences. After instrumentation, injury on neighbouring teeth was assessed. Instrumented implants and non-instrumented controls were subjected to 1.2 million cycles of thermo-mechanical loading in a chewing machine. Afterwards, maximum fracture load for all implants and an additional five pristine control implants was tested.
Results

Generally, damage of neighbour teeth was a frequent finding (33 ± 56% of all cases) with considerable inter-group differences. No considerable inter-group difference for the residual implant thickness was found for different areas assessed. No implant fractured during cyclic loading. Fracture load was reduced after cyclic loading of uninstrumented implants from 2,724 ± 70 N to 2,299 ± 127 N, and after implantoplasty to 1,737 ± 165 N, while no effect by the instrumentation sequence could be observed.
Conclusions

Both implantoplasty and cyclic loading were shown to reduce the implants' maximum bending strength. Cyclic loading in a laboratory masticator, simulating a five-year equivalent of chewing, did not result in fractured implants. Since neighbouring tooth injury was assessed often, care should be taken with the selection of suitable instruments.

Additional indexing

Item Type:Dissertation (monographical)
Referees:Sahrmann Philipp, Schmidlin Patrick
Communities & Collections:04 Faculty of Medicine > Center for Dental Medicine > Clinic of Conservative and Preventive Dentistry
UZH Dissertations
Dewey Decimal Classification:610 Medicine & health
Language:English
Place of Publication:Zürich
Date:2021
Deposited On:04 Feb 2022 15:50
Last Modified:05 Feb 2022 04:27
OA Status:Closed
Related URLs:https://uzb.swisscovery.slsp.ch/permalink/41SLSP_UZB/1d8t6qj/alma99117137752405508 (Library Catalogue)
https://www.zora.uzh.ch/id/eprint/207215/
Full text not available from this repository.