Flicker electroretinogram in newborn infants

Hanson, James V M; Weber, Caroline; Pfäffli, Oliver Andreas; Bassler, Dirk; McCulloch, Daphne L; Gerth-Kahlert, Christina (2022). Flicker electroretinogram in newborn infants. Documenta ophthalmologica. Advances in ophthalmology, 145(3):175-184.

Abstract

PURPOSE: To develop and validate a flicker electroretinogram (ERG) protocol in term-born neonates as a potential tool for assessing preterm infants at risk of developing retinopathy of prematurity.

METHODS: A custom flicker ERG protocol was developed for use with the hand-held RETeval® electrophysiology device. Feasibility of measuring flicker ERG through closed eyelids and without mydriasis was established in a pilot study enabling optimisation of the test protocol. Following this, healthy term-born neonates (gestational age 37-42 weeks) were recruited at the Neonatology clinic of the University Hospital Zurich. Flicker ERG recordings were performed using proprietary disposable skin electrodes during the first four days of life when the infants were sleeping. Flicker stimuli were presented at 28.3 Hz for a stimulus series at 3, 6, 12, 30, and 50 cd·s/m$^{2}$, with two measurements at each stimulus level. Results were analysed offline. Flicker ERG peak times and amplitudes were derived from the averaged measurements per stimulus level for each subject.

RESULTS: 28 term-born neonates were included in the analysis. All infants tolerated the testing procedure well. Flicker ERG recording was achieved in all subjects with reproducible flicker ERG waveforms for 30 and 50 cd·s/m$^{2}$ stimuli. Reproducible ERGs were recorded in the majority of infants for the weaker stimuli (with detectable ERGs in 20/28, 25/28, and 27/28 at 3, 6, and 12 cd·s/m$^{2}$, respectively). Flicker ERG amplitudes increased with increasing stimulus strength, with peak times concurrently decreasing slightly.

CONCLUSION: Flicker ERG recording is feasible and reliably recorded in sleeping neonates through closed eyelids using skin electrodes and without mydriasis. Flicker ERG amplitude decreases for lower luminance flicker but remains detectable for 3 cd·s/m$^{2}$ flicker in the majority of healthy term-born neonates. These data provide a basis to study retinal function in premature infants using this protocol.

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Ophthalmology Clinic
04 Faculty of Medicine > University Hospital Zurich > Clinic for Neonatology
Dewey Decimal Classification:610 Medicine & health
Scopus Subject Areas:Health Sciences > Ophthalmology
Life Sciences > Sensory Systems
Health Sciences > Physiology (medical)
Language:English
Date:December 2022
Deposited On:17 Nov 2022 12:15
Last Modified:20 Mar 2025 04:37
Publisher:Springer
ISSN:0012-4486
OA Status:Hybrid
Free access at:Publisher DOI. An embargo period may apply.
Publisher DOI:https://doi.org/10.1007/s10633-022-09889-5
Related URLs:https://www.zora.uzh.ch/id/eprint/251511/
PubMed ID:36199003
Download PDF  'Flicker electroretinogram in newborn infants'.
Preview
  • Content: Published Version
  • Language: English
  • Licence: Creative Commons: Attribution 4.0 International (CC BY 4.0)

Metadata Export

Statistics

Citations

Dimensions.ai Metrics
3 citations in Web of Science®
3 citations in Scopus®
Google Scholar™

Altmetrics

Downloads

11 downloads since deposited on 17 Nov 2022
4 downloads since 12 months
Detailed statistics

Authors, Affiliations, Collaborations

Similar Publications