A novel reference model for dental scanning system evaluation: analysis of five intraoral scanners


Karakas-Stupar, Irina; Zitzmann, Nicola Ursula; Joda, Tim (2022). A novel reference model for dental scanning system evaluation: analysis of five intraoral scanners. The Journal of Advanced Prosthodontics, 14(2):63.

Abstract

Purpose: The aim of this in vitro study was to investigate the accuracy (trueness and precision) of five intraoral scanners (IOS) using a novel reference model for standardized performance evaluation.

Materials and methods: Five IOSs (Medit i500, Omnicam, Primescan, Trios 3, Trios 4) were used to digitize the reference model, which represented a simplified full-arch situation with four abutment teeth. Each IOS was used five times by an experienced operator, resulting in 25 STL (Standard Tessellation Language) files. STL data were imported into 3D software (Final Surface®) and examined for inter- and intra-group analyses. Deviations in the parameter matching error were calculated. ANOVA F-test and Kruskal-Wallis test were applied for inter-group comparisons (α = .05); and the coefficient of variation (CV) was calculated for intra-group comparisons (in % ± SD).

Results: Primescan (matching error value: 0.015), Trios 3 (0.016), and Trios 4 (0.018) revealed comparable results with significantly higher accuracy compared to Medit i500 (0.035) and Omnicam (0.028) (P < .001). For intra-group comparison, Trios 4 demonstrated the most homogenous results (CV 15.8%).

Conclusion: The novel reference model investigated in this study can be used to assess the performance of dental scanning technologies in the daily routine setting and in research settings.

Keywords: Fixed prosthodontics; Intraoral optical scanning (IOS); Precision; Reference model; Trueness.

Abstract

Purpose: The aim of this in vitro study was to investigate the accuracy (trueness and precision) of five intraoral scanners (IOS) using a novel reference model for standardized performance evaluation.

Materials and methods: Five IOSs (Medit i500, Omnicam, Primescan, Trios 3, Trios 4) were used to digitize the reference model, which represented a simplified full-arch situation with four abutment teeth. Each IOS was used five times by an experienced operator, resulting in 25 STL (Standard Tessellation Language) files. STL data were imported into 3D software (Final Surface®) and examined for inter- and intra-group analyses. Deviations in the parameter matching error were calculated. ANOVA F-test and Kruskal-Wallis test were applied for inter-group comparisons (α = .05); and the coefficient of variation (CV) was calculated for intra-group comparisons (in % ± SD).

Results: Primescan (matching error value: 0.015), Trios 3 (0.016), and Trios 4 (0.018) revealed comparable results with significantly higher accuracy compared to Medit i500 (0.035) and Omnicam (0.028) (P < .001). For intra-group comparison, Trios 4 demonstrated the most homogenous results (CV 15.8%).

Conclusion: The novel reference model investigated in this study can be used to assess the performance of dental scanning technologies in the daily routine setting and in research settings.

Keywords: Fixed prosthodontics; Intraoral optical scanning (IOS); Precision; Reference model; Trueness.

Statistics

Citations

Dimensions.ai Metrics
2 citations in Web of Science®
2 citations in Scopus®
Google Scholar™

Altmetrics

Downloads

12 downloads since deposited on 27 Jan 2023
9 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > Center for Dental Medicine > Clinic of Reconstructive Dentistry
Dewey Decimal Classification:610 Medicine & health
Scopus Subject Areas:Health Sciences > Oral Surgery
Health Sciences > Dentistry (miscellaneous)
Language:English
Date:April 2022
Deposited On:27 Jan 2023 10:35
Last Modified:28 Feb 2024 02:47
Publisher:Korean Academy
ISSN:2005-7806
OA Status:Gold
Free access at:Publisher DOI. An embargo period may apply.
Publisher DOI:https://doi.org/10.4047/jap.2022.14.2.63
PubMed ID:35601353