Mitacchione, Gianfranco; Schiavone, Marco; Gasperetti, Alessio; Arabia, Gianmarco; Breitenstein, Alexander; et al; Hofer, Daniel; Saguner, Ardan M (2023). Outcomes of leadless pacemaker implantation following transvenous lead extraction in high-volume referral centers: Real-world data from a large international registry. Heart Rhythm, 20(3):395-404.

Abstract

BACKGROUND

Limited data on the real-world safety and efficacy of leadless pacemakers (LPMs) post-transvenous lead extraction (TLE) are available.

OBJECTIVE

The purpose of this study was to assess the long-term safety and effectiveness of LPMs following TLE in comparison with LPMs de novo implantation.

METHODS

Consecutive patients who underwent LPM implantation in 12 European centers joining the International LEAdless PacemakEr Registry were enrolled. The primary end point was the comparison of LPM-related complication rate at implantation and during follow-up (FU) between groups. Differences in electrical performance were deemed secondary outcomes.

RESULTS

Of the 1179 patients enrolled, 15.6% underwent a previous TLE. During a median FU of 33 (interquartile range 24-47) months, LPM-related major complications and all-cause mortality did not differ between groups (TLE group: 1.6% and 5.4% vs de novo group: 2.2% and 7.8%; P = .785 and P = .288, respectively). Pacing threshold (PT) was higher in the TLE group at implantation and during FU, with very high PT (>2 V@0.24 ms) patients being more represented than in the de novo implantation group (5.4% vs 1.6 %; P = .004). When the LPM was deployed at a different right ventricular (RV) location than the one where the previous transvenous RV lead was extracted, a lower proportion of high PT (>1-2 V@0.24 ms) patients at implantation, 1-month FU, and 12-month FU (5.9% vs 18.2%, P = .012; 3.4% vs 12.9%, P = .026; and 4.3% vs 14.5%, P = .037, respectively) was found.

CONCLUSION

LPMs showed a satisfactory safety and efficacy profile after TLE. Better electrical parameters were obtained when LPMs were implanted at a different RV location than the one where the previous transvenous RV lead was extracted.

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Clinic for Cardiology
Dewey Decimal Classification:610 Medicine & health
Scopus Subject Areas:Health Sciences > Cardiology and Cardiovascular Medicine
Health Sciences > Physiology (medical)
Language:English
Date:1 March 2023
Deposited On:08 Feb 2023 14:09
Last Modified:27 Apr 2025 01:36
Publisher:Elsevier
ISSN:1547-5271
OA Status:Closed
Publisher DOI:https://doi.org/10.1016/j.hrthm.2022.12.002
PubMed ID:36496135
Full text not available from this repository.

Metadata Export

Statistics

Citations

Dimensions.ai Metrics
26 citations in Web of Science®
25 citations in Scopus®
Google Scholar™

Altmetrics

Authors, Affiliations, Collaborations

Similar Publications