In vitro evaluation of how the presence of the stent retriever and microcatheter influences aspiration parameters in thrombectomy according to their position inside the aspiration catheter

Bernava, Gianmarco; Brina, Olivier; Reymond, Philippe; Rosi, Andrea; Hofmeister, Jeremy; Yilmaz, Hasan; Muster, Michel; Kulcsar, Zsolt; Lovblad, Karl-Olof; Machi, Paolo (2024). In vitro evaluation of how the presence of the stent retriever and microcatheter influences aspiration parameters in thrombectomy according to their position inside the aspiration catheter. Interventional Neuroradiology, 30(4):489-495.

Abstract

BACKGROUND: Several variations of the combined thrombectomy technique for acute ischemic stroke using a stent retriever and aspiration catheter have been described. The aim of our study was to assess how the presence of the microcatheter and stent retriever affect the basic aspiration parameters, namely, flow rate and aspiration force, depending on their position within the aspiration catheter.

METHODS: Two experimental set-ups were designed to assess changes in flow rate and aspiration force according to the position of the stent retriever and microcatheter within the aspiration catheter.

RESULTS: The transition of the stent retriever and microcatheter from the distal to proximal position resulted in a progressive increase in the flow rate, but with no impact on aspiration force. Additionally, the size of the stent retriever had no significant effect on flow rate changes and the reduction in flow rate was related to the microcatheter diameter. Negative pressure generated inside the aspiration catheter impacted on its distal segment located beyond the radiopaque marker, thus leading to its partial collapse. As a consequence, the measured aspiration force was lower than the theoretical aspiration force level for all tested aspiration catheters.

CONCLUSIONS: In our experimental model, the position of the stent retriever and microcatheter within the aspirator catheter affected the flow rate, but not the aspiration force. Negative pressure generated within the aspiration catheter appeared to determine a partial collapse of the distal segment that resulted in a less effective aspiration force than the theoretical aspiration force level.

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Clinic for Neuroradiology
Dewey Decimal Classification:610 Medicine & health
Scopus Subject Areas:Health Sciences > Radiology, Nuclear Medicine and Imaging
Health Sciences > Neurology (clinical)
Health Sciences > Cardiology and Cardiovascular Medicine
Language:German
Date:August 2024
Deposited On:04 Apr 2023 07:16
Last Modified:26 Feb 2025 02:37
Publisher:Edizioni del Centauro
ISSN:1591-0199
OA Status:Hybrid
Free access at:Publisher DOI. An embargo period may apply.
Publisher DOI:https://doi.org/10.1177/15910199221135040
PubMed ID:36348632
Download PDF  'In vitro evaluation of how the presence of the stent retriever and microcatheter influences aspiration parameters in thrombectomy according to their position inside the aspiration catheter'.
Preview
  • Content: Published Version
  • Language: English
  • Licence: Creative Commons: Attribution 4.0 International (CC BY 4.0)

Metadata Export

Statistics

Citations

Dimensions.ai Metrics
2 citations in Web of Science®
3 citations in Scopus®
Google Scholar™

Altmetrics

Downloads

45 downloads since deposited on 04 Apr 2023
30 downloads since 12 months
Detailed statistics

Authors, Affiliations, Collaborations

Similar Publications