Evaluation of the i-STAT portable clinical analyzer in chickens (Gallus gallus)


Steinmetz, H W; Vogt, Rainer; Kästner, Sabine Beate Rita; Riond, Barbara; Hatt, Jean-Michel (2007). Evaluation of the i-STAT portable clinical analyzer in chickens (Gallus gallus). Journal of Veterinary Diagnostic Investigation, 19(4):382-388.

Abstract

The i-STAT portable clinical analyzer (PCA) was evaluated for performance in avian species. With the EG7+ cartridge, which provided results for hydrogen ion concentration, oxygen tension, carbon dioxide tension, sodium, potassium, ionized calcium, hematocrit, and various calculated parameters, analytical accuracy and precision were tested by comparing obtained values to those of established traditional blood gas and chemistry analyzers. Deming's regression and bias plots were used to compare i-STAT results with those obtained by laboratory professionals using benchtop analyzers. The reliability of the i-STAT PCA with EG7+ cartridges was good, with 0-5.7% system failures in measured values. Regression statistics were good for all blood gas analytes and acceptable for electrolytes and calculated parameters, except for potassium and base excess, for which the regression data or the discrepancy between the methods was too large. The system was reliable and easy to use and had an overall acceptable accuracy in avian species. These features, together with portability and small required blood volumes, make the i-STAT suitable for point-of-care use in critical avian patients, although single values require careful interpretation.

Abstract

The i-STAT portable clinical analyzer (PCA) was evaluated for performance in avian species. With the EG7+ cartridge, which provided results for hydrogen ion concentration, oxygen tension, carbon dioxide tension, sodium, potassium, ionized calcium, hematocrit, and various calculated parameters, analytical accuracy and precision were tested by comparing obtained values to those of established traditional blood gas and chemistry analyzers. Deming's regression and bias plots were used to compare i-STAT results with those obtained by laboratory professionals using benchtop analyzers. The reliability of the i-STAT PCA with EG7+ cartridges was good, with 0-5.7% system failures in measured values. Regression statistics were good for all blood gas analytes and acceptable for electrolytes and calculated parameters, except for potassium and base excess, for which the regression data or the discrepancy between the methods was too large. The system was reliable and easy to use and had an overall acceptable accuracy in avian species. These features, together with portability and small required blood volumes, make the i-STAT suitable for point-of-care use in critical avian patients, although single values require careful interpretation.

Statistics

Citations

Dimensions.ai Metrics
46 citations in Web of Science®
50 citations in Scopus®
Google Scholar™

Altmetrics

Downloads

120 downloads since deposited on 15 Apr 2008
10 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:05 Vetsuisse Faculty > Veterinary Clinic > Department of Small Animals
Dewey Decimal Classification:570 Life sciences; biology
630 Agriculture
Scopus Subject Areas:Health Sciences > General Veterinary
Uncontrolled Keywords:Avian, blood gas, electrolytes, emergency, point-of-care
Language:English
Date:2007
Deposited On:15 Apr 2008 14:35
Last Modified:24 Jun 2022 19:18
Publisher:American Association of Veterinary Laboratory Diagnosticians
ISSN:1040-6387
OA Status:Hybrid
Publisher DOI:https://doi.org/10.1177/104063870701900407
PubMed ID:17609347
  • Content: Published Version
  • Language: English