Navigation auf zora.uzh.ch

Search ZORA

ZORA (Zurich Open Repository and Archive)

Estimated Radiation Dose to the Lens During Endourologic Procedures: The Role of Leaded Glasses and the ALARA Protocol-An ESU/ESUT-YAU Endourology Group Collaboration

Bravo-Balado, Alejandra; Fontanet, Sofia; Skolarikos, Andreas; Gozen, Ali Serdar; Somani, Bhaskar K; Traxer, Olivier; Papatsoris, Athanasios; Ruiz Martínez, Agustín; Keller, Etienne Xavier; Pietropaolo, Amelia; Tonyali, Senol; Tailly, Thomas; Esperto, Francesco; Liatsikos, Evangelos; Kanashiro, Andrés K; Angerri, Oriol; Emiliani, Esteban (2023). Estimated Radiation Dose to the Lens During Endourologic Procedures: The Role of Leaded Glasses and the ALARA Protocol-An ESU/ESUT-YAU Endourology Group Collaboration. Journal of Endourology, 37(8):935-939.

Abstract

Purpose: On February 6, 2018, the European Atomic Energy Community reduced the annual equivalent dose limit for the lens from 150 to 20 mSv/year, because of its association with cataracts at low radiation doses. Our aim was to estimate the radiation doses received by the lens during endourologic procedures that require fluoroscopy. Materials and Methods: Multicenter study including prospective data of annual eye dosimeters between 2017 and 2020. Four endourologists used an eye dosimeter in endourologic procedures that require fluoroscopy (ureteroscopy, retrograde intrarenal surgery, and percutaneous nephrolithotomy). Surgeons 1 and 2 wore leaded glasses; surgeon 1 also used the as low as reasonably achievable (ALARA) protocol. Descriptive statistical analysis using SPSS 25.0 was conducted.
Results: Surgeons 1, 2, 3, and 4 performed a median of 159, 586, 102, and 129 endourologic procedures per year, respectively, for a total of 641, 2340, 413, and 350 procedures between 2017 and 2020. The median annual dose of lens radiation exposure was 0.16, 1.18, 3.79, and 1.42 mSv per year, respectively, which corresponds to 0.001, 0.009, 0.024, and 0.012 mSv per procedure. The two surgeons who used leaded glasses registered a lower radiation dose per procedure (0.001 vs 0.027). Similarly, the urologist who used the ALARA protocol registered the lowest lens radiation dose compared with the three surgeons who did not use it (0.001 vs 0.023).
Conclusions: The endourologists who participated in this study effectively comply with current guidelines on radiation exposure to the lens. Registered eye lens radiation does not seem to be related to the number of procedures but rather to the use of leaded glasses and the ALARA protocol.

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Urological Clinic
Dewey Decimal Classification:610 Medicine & health
Scopus Subject Areas:Health Sciences > Urology
Language:English
Date:August 2023
Deposited On:07 Dec 2023 08:55
Last Modified:30 Dec 2024 02:52
Publisher:Mary Ann Liebert
ISSN:0892-7790
OA Status:Closed
Publisher DOI:https://doi.org/10.1089/end.2023.0061
PubMed ID:37337653

Metadata Export

Statistics

Citations

Dimensions.ai Metrics
1 citation in Web of Science®
1 citation in Scopus®
Google Scholar™

Altmetrics

Downloads

1 download since deposited on 07 Dec 2023
0 downloads since 12 months
Detailed statistics

Authors, Affiliations, Collaborations

Similar Publications