Characterization of spatial integrity with active and passive implants in a low-field magnetic resonance linear accelerator scanner

Pouymayou, Bertrand; Perez-Haas, Yoel; Allemann, Florin; Saguner, Ardan M; Andratschke, Nicolaus; Guckenberger, Matthias; Tanadini-Lang, Stephanie; Wilke, Lotte (2024). Characterization of spatial integrity with active and passive implants in a low-field magnetic resonance linear accelerator scanner. Physics and Imaging in Radiation Oncology, 30:100576.

Abstract

BACKGROUND AND PURPOSE
Standard imaging protocols can guarantee the spatial integrity of magnetic resonance (MR) images utilized in radiotherapy. However, the presence of metallic implants can significantly compromise this integrity. Our proposed method aims at characterizing the geometric distortions induced by both passive and active implants commonly encountered in planning images obtained from a low-field 0.35 T MR-linear accelerator (LINAC).

MATERIALS AND METHODS
We designed a spatial integrity phantom defining 1276 control points and covering a field of view of 20x20x20 cm3. This phantom was scanned in a water tank with and without different implants used in hip and shoulder arthroplasty procedures as well as with active cardiac stimulators. The images were acquired with the clinical planning sequence (balanced steady-state free-precession, resolution 1.5x1.5x1.5 mm3). Spatial integrity was assessed by the Euclidian distance between the control point detected on the image and their theoretical locations. A first plane free of artefact (FPFA) was defined to evaluate the spatial integrity beyond the larger banding artefact.

RESULTS
In the region extending up to 20 mm from the largest banding artefacts, the tested passive and active implants could cause distortions up to 2 mm and 3 mm, respectively. Beyond this region the spatial integrity was recovered and the image could be considered as unaffected by the implants.

CONCLUSIONS
We characterized the impact of common implants on a low field MR-LINAC planning sequence. These measurements could support the creation of extra margin while contouring organs at risk and target volumes in the vicinity of implants.

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Department of Trauma Surgery
04 Faculty of Medicine > University Hospital Zurich > Clinic for Radiation Oncology
Dewey Decimal Classification:610 Medicine & health
Scopus Subject Areas:Physical Sciences > Radiation
Health Sciences > Oncology
Health Sciences > Radiology, Nuclear Medicine and Imaging
Language:English
Date:18 April 2024
Deposited On:06 May 2024 14:47
Last Modified:31 Dec 2024 02:37
Publisher:Elsevier
ISSN:2405-6316
OA Status:Gold
Free access at:Publisher DOI. An embargo period may apply.
Publisher DOI:https://doi.org/10.1016/j.phro.2024.100576
PubMed ID:38644933
Download PDF  'Characterization of spatial integrity with active and passive implants in a low-field magnetic resonance linear accelerator scanner'.
Preview
  • Content: Published Version
  • Language: English
  • Licence: Creative Commons: Attribution 4.0 International (CC BY 4.0)

Metadata Export

Statistics

Citations

Altmetrics

Downloads

6 downloads since deposited on 06 May 2024
6 downloads since 12 months
Detailed statistics

Authors, Affiliations, Collaborations

Similar Publications