Effect of vertical stopcock position on start-up fluid delivery in syringe pumps used for microinfusions

Weiss, Markus; Wendel-Garcia, Pedro David; Grass, Beate; Kleine-Brueggeney, Maren (2024). Effect of vertical stopcock position on start-up fluid delivery in syringe pumps used for microinfusions. Journal of Clinical Monitoring and Computing, 38(5):1209-1218.

Abstract

The purpose of this in vitro study was to evaluate the impact of the vertical level of the stopcock connecting the infusion line to the central venous catheter on start-up fluid delivery in microinfusions. Start-up fluid delivery was measured under standardized conditions with the syringe outlet and liquid flow sensors positioned at heart level (0 cm) and exposed to a simulated CVP of 10 mmHg at a set flow rate of 1 ml/h. Flow and intraluminal pressures were measured with the infusion line connected to the stopcock primarily placed at vertical levels of 0 cm, + 30 cm and − 30 cm or primarily placed at 0 cm and secondarily, after connecting the infusion line, displaced to + 30 cm and − 30 cm. Start-up fluid delivery 10 s after opening the stopcock placed at zero level and after opening the stopcock primarily connected at zero level and secondary displaced to vertical levels of + 30 cm and – 30 cm were similar (− 10.52 [− 13.85 to − 7.19] µL; − 8.84 [− 12.34 to − 5.33] µL and − 11.19 [− 13.71 to − 8.67] µL (p = 0.469)). Fluid delivered at 360 s related to 65% (zero level), 71% (+ 30 cm) and 67% (− 30 cm) of calculated infusion volume (p = 0.395). Start-up fluid delivery with the stopcock primarily placed at + 30 cm and − 30 cm resulted in large anterograde and retrograde fluid volumes of 34.39 [33.43 to 35.34] µL and − 24.90 [− 27.79 to − 22.01] µL at 10 s, respectively (p < 0.0001). Fluid delivered with the stopcock primarily placed at + 30 cm and − 30 cm resulted in 140% and 35% of calculated volume at 360 s, respectively (p < 0.0001). Syringe infusion pumps should ideally be connected to the stopcock positioned at heart level in order to minimize the amounts of anterograde and retrograde fluid volumes after opening of the stopcock.

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Institute of Intensive Care Medicine
04 Faculty of Medicine > University Hospital Zurich > Clinic for Neonatology
Dewey Decimal Classification:610 Medicine & health
Scopus Subject Areas:Health Sciences > Health Informatics
Health Sciences > Critical Care and Intensive Care Medicine
Health Sciences > Anesthesiology and Pain Medicine
Language:English
Date:October 2024
Deposited On:06 May 2024 15:54
Last Modified:27 Feb 2025 02:44
Publisher:Springer
ISSN:1387-1307
OA Status:Hybrid
Free access at:Publisher DOI. An embargo period may apply.
Publisher DOI:https://doi.org/10.1007/s10877-024-01156-z
PubMed ID:38619717
Project Information:
  • Funder: University of Zurich
  • Grant ID:
  • Project Title:
Download PDF  'Effect of vertical stopcock position on start-up fluid delivery in syringe pumps used for microinfusions'.
Preview
  • Content: Published Version
  • Language: English
  • Licence: Creative Commons: Attribution 4.0 International (CC BY 4.0)

Metadata Export

Statistics

Citations

Dimensions.ai Metrics
1 citation in Web of Science®
1 citation in Scopus®
Google Scholar™

Altmetrics

Downloads

3 downloads since deposited on 06 May 2024
3 downloads since 12 months
Detailed statistics

Authors, Affiliations, Collaborations

Similar Publications