Navigation auf zora.uzh.ch

Search

ZORA (Zurich Open Repository and Archive)

Ultrafast nuclear myocardial perfusion imaging on a new gamma camera with semiconductor detector technique: first clinical validation

Buechel, R R; Herzog, B A; Husmann, L; Burger, I A; Pazhenkottil, A P; Treyer, V; Valenta, I; von Schulthess, P; Nkoulou, R; Wyss, C A; Kaufmann, P A (2010). Ultrafast nuclear myocardial perfusion imaging on a new gamma camera with semiconductor detector technique: first clinical validation. European Journal of Nuclear Medicine and Molecular Imaging, 37(4):773-778.

Abstract

PURPOSE: To assess the diagnostic performance of a novel ultrafast cardiac gamma camera with cadmium-zinc-telluride (CZT) solid-state semiconductor detectors for nuclear myocardial perfusion imaging (MPI). METHODS: The study group comprised 75 consecutive patients (55 men, BMI range 19-45 kg/m(2)) who underwent a 1-day (99m)Tc-tetrofosmin adenosine-stress/rest imaging protocol. Scanning was performed first on a conventional dual-detector SPECT gamma camera (Ventri, GE Healthcare) with a 15-min acquisition time each for stress and rest. All scans were immediately repeated on an ultrafast CZT camera (Discovery 530 NMc, GE Healthcare) with a 3-min scan time for stress and a 2-min scan time for rest. Clinical agreement (normal, ischaemia, scar) between CZT and SPECT was assessed for each patient and for each coronary territory using SPECT MPI as the reference standard. Segmental myocardial tracer uptake values (percent of maximum) using a 20-segment model and left ventricular ejection fraction (EF) values obtained using CZT were compared with those obtained using conventional SPECT by intraclass correlation and by calculating Bland-Altman limits of agreement. RESULTS: There was excellent clinical agreement between CZT and conventional SPECT on a per-patient basis (96.0%) and on a per-vessel territory basis (96.4%) as shown by a highly significant correlation between segmental tracer uptake values (r=0.901, p<0.001). Similarly, EF values for both scanners were highly correlated (r=0.976, p<0.001) with narrow Bland-Altman limits of agreement (-5.5-10.6%). CONCLUSION: The novel CZT camera allows a more than fivefold reduction in scan time and provides clinical information equivalent to conventional standard SPECT MPI.

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > Zurich Center for Integrative Human Physiology (ZIHP)
04 Faculty of Medicine > University Hospital Zurich > Clinic for Nuclear Medicine
04 Faculty of Medicine > University Hospital Zurich > Clinic for Cardiology
Dewey Decimal Classification:570 Life sciences; biology
610 Medicine & health
Scopus Subject Areas:Health Sciences > Radiology, Nuclear Medicine and Imaging
Language:English
Date:April 2010
Deposited On:12 Nov 2010 15:25
Last Modified:04 Sep 2024 01:39
Publisher:Springer
ISSN:1619-7070
OA Status:Green
Publisher DOI:https://doi.org/10.1007/s00259-009-1375-7
PubMed ID:20107783

Metadata Export

Statistics

Citations

Dimensions.ai Metrics
144 citations in Web of Science®
158 citations in Scopus®
Google Scholar™

Altmetrics

Downloads

125 downloads since deposited on 12 Nov 2010
29 downloads since 12 months
Detailed statistics

Authors, Affiliations, Collaborations

Similar Publications